Push Start when the calibration weights are on, and the hopper weight is stable.
The weight is sampled. This requires, again, that the weight value is stable.
This screen indicates that most likely, no calibration weights were applied. Your only
option is to push
Cancel
.
Normally, after the weight has been sampled, the changes that are about to be applied
are displayed. You have the option of pushing
Accept
, to apply the changes, or
Cancel
to have the controller ignore the entire calibration procedure.
NOTE:
The logical input
Allow Calibr
has to be ON for the
Accept
button to appear.
Electronic Calibration
The Electronic (E-Cal) Procedure is provided as an alternate method to set the
Scale
Factor
. It is used when it is not possible to run either a Material or Weight procedure. The
HPAD board can inject a known artificial load cell signal to simulate weight. You must
calculate a parameter, called
, see page 28. Use the following formula to
calculate the E-Cal Live Load mV/V:
Cells
Load
of
Capacity
Cell
Load
of
rating
mV/V
*
Weight
Design
=
mV/V
ECAL
35.00.EX O&M Manual
50