If there are more than one load cells supporting the hopper, add the load cells capacities.
For example, for three 1000 lb load cells:
Kg
1361
2.204
3
*
1000
=
Cells
Load
of
Capacity
=
If the load cells are 3 mV/V, and the Design Weight is 400 kg, then:
mV/V
0.8817
1361
3
*
400
=
mV/V
ECAL
=
Enter ECAL Liveload
, in Calibrate Settings, before running the procedure. The hopper
must be empty, and a Zero Procedure must be performed immediately before the E-Cal
Procedure.
To start the test, push the
Ecal
button from the Calibration menu.
If the
ECAL Live Load
value is incorrect or if you do not wish to complete the E-Cal
Procedure for any reason, push the
Cancel
button to return to the Calibration menu. If
you are ready to execute the E-Cal Procedure, push the
Start
button.
The procedure goes through two phases. In the first phase, the HPAD sets up a known
artificial load cell signal, in mV/V.
In the second phase, the artificial weight indication in sampled.
35.00.EX O&M Manual
51