51
Error type
Possible causes
Troubleshooting
Controls within the reagent
range.
Unexpected result of patient
samples.
Analytical error
Wrong incubation time
Follow the manufacturer’s
instructions.
Analytical error
Wrong test sequence.
Follow the manufacturer’s
instructions.
Irregular results within the
test.
Controls may be within or out
of the reagent range.
Analytical error
Imprecise manual pipetting of
sample and reagent.
The pipette has to be main-
tained. The as accessory
available automatic pipette of
the MC 10
plus
is delivered with
manual. Please practise the
pipetting technique. The
instructions for the correct
pipetting technique are in
chapter 4 (pipetting).
Wrong dispensing position: it is
very important that the reagent
is always dispensed from the
same position. Please find the
instructions for the correct
pipetting technique in chapter 4
(pipetting).
Reagent in particle form has not
been mixed before
employment. Close the opening
of the tube with a cap or with
Parafilm™, turn round the tube
and mix it gently.
Sample and first reagent have
not been mixed. After sample
and reagent have been
dispensed take the cuvette out
of the pre-heating station and
sway it gently 5 or 6 times for
dispensing the mixture
constantly on the bottom of the
cuvette.