20
3.3
Test parameters
After switching on the analyser select the menu test parameter in the start menu by using the arrow
keys and confirm your selection by pressing ENTER. (The test parameter menu is protected by a PIN and
can only be entered by instructed staff).
Start menu
Measure mode
Test parameter
Settings menu
Patient memory
Service menu
Here different parameter specifications can be determined. By selecting the parameter to be modified
via the arrow keys and by confirming with ENTER you get into the display of the according parameter.
Depending on the parameter following settings can be done:
PT
PTT
FIB
TZ
FAK
Test name
PT, NT or TT
PTT
FIB
TZ
FAK
Incubation time
30, 40, 50, 60,...
Result format
number of digits before and after decimal point
Calibration
several values
1)
-
several values
1)
-
sev. values
1)
Calculation
lin/rezi
-
log/log
log/lin
Result unit
%
sec.
g/L.
sec.
sec.
Calibration value
max + min
-
max + min
-
max + min
INR-Calculation
YES / NO
-
-
-
-
INR-Std.-Value
value
2)
-
-
-
-
ISI-Value
value
3)
-
-
-
-
Ball or optical method
ball (mechanical)
Single or double
single / double
Deviation (VC)
freely adjustable
Start delay
freely adjustable between 3,5 and 30 seconds
Timeout time
freely adjustable (depending on result format)
FP 1
FP 2
FP 3
Emergency Photometer
Test name
freely programmable
this
parameter
reverts to
two other
parameters
this
parameter
reverts to one
other
parameter
Incubation time
30, 40, 50, 60,...
Result format
number of digits before and after decimal point
Calibration
several values
1)
Calculation
several calculation formulas
Result unitt
freely programmable
Calibration value
max + min
INR-calculation
YES / NO
INR-Std.
value
2)
ISI-value
value
3)
Ball or optical method
ball (mechanical) or optical
Single or double
single / double
double
Deviation (VC)
freely adjustable
(6%)
Start delay
freely adjustable between 3,5 and 30 seconds
Timeout time
freely adjustable (depending on result format)