24
Chromogenic:
After a freely selectable interval (10 / 20 / 30 sec. or 20 / 40 / 60 sec. or 30 / 60 / 90 sec.) the
extinction modifications are measured. The modification is calculated in mOD/min. The
concentration is calculated by comparison with a standard curve.
Delta E:
The first test value is established after the end of the start delay time as for Clot 1. The second
value is established after the timeout time. The modification is displayed in mOD. The
concentration is calculated by comparison with a standard curve.
Test points of an optical clotting measurement
(measurement by 30-60-90 seconds)
0
50
100
150
200
0
20
40
60
80
100
120
Time in sec.
T
e
st
v
a
lu
e
i
n
m
O
D
1st test
2nd test
3rd test
Example for chromogenic measurement
Example for time measurement Delta E
Test points of an optical clotting measurement
(measurement by defined time)
0
50
100
150
200
0
20
40
60
80
100
120
Time in sec.
T
es
t
va
lu
e
in
m
O
D
defined test time
Test value