![Merlin medical MC10 PLUS Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/merlin-medical/mc10-plus/mc10-plus_operation-manual_1774758033.webp)
30
Print by patient-ID:
With this menu item the patient results can be printed.
Before the printout the patient-ID can be entered via the numerical keys, counted up and down with the
arrow keys
and
or read in with the barcode scanner. Then the print command can be given with the
ENTER-key
.
Print by date:
All results of all patients of a defined date are printed.
Send online by patient-ID:
With this function you can send the patient date to the connected PC.
Send online by date:
With this function you can send the data of a prior defined date to the connected PC.
Delete:
With the delete function the memory of the instrument is deleted completely (all test results). The
moment of deletion depends on the internal laboratory organisation. It is recommended to carry out
the deletion procedure every morning before start of work.
Altogether up to 1,000 results can be stored whereby the storing procedure is basing on the FIFO-
principle (first-in-first-out). When result no. 1,001 is stored the result no. 1 is deleted automatically from
the memory.
3.7
Service-menu
After switching on the analyser select the menu service menu in the start menu by using the arrow keys
and confirm your selection by pressing ENTER. (The service menu is protected by a PIN and can only be
entered by instructed staff).
In this menu you can - with the support of the manufacturer - you can modify instrument specific
parameters as e.g. measuring block temperature, measuring channel turning speed etc. The service
menu can only be opened after prior consultation with the manufacturer.