Location of the instrument
1.
Place the MC 10
plus
on a plane, stable, vibration- and dust-free work surface which is deep and
wide enough to ensure the air circulation of the instrument. For ensuring a sufficient cooling of
the analyser the distance between instrument and wall respectively to another object has to be
at least 10 cm. The instrument should not be placed next to centrifuges or other instruments
which could cause vibrations.
Minimum space requirements:
-
width
79 cm
(width of the instrument:
59 cm)
-
depth
53 cm
(depth of the instrument:
43 cm)
2.
Position the MC 10
plus
in an area with low humidity and little variations in temperature. The
device should not be placed directly under ventilation shafts which cause strong draughts.
3.
Place the MC 10
plus
in an area which is protected from direct sun light.
4.
The distance between the analyser and the socket may not exceed 3 m. Other instruments with
high power consumption and which are frequently switched on and off as e.g. centrifuges, air
conditionings or refrigerators should not be connected to the same circuit. When switching on
and off such instruments the voltage drop can be strong enough to have a negative effect on the
proper operation of the MC 10
plus
.
Attention!
If the user is electrified a discharge may happen at the MC 10
plus
. This
discharge has no influence on the function of the MC 10
plus
.