14
2.5
Connection demands
1.
Before the electrical installation is carried out it has to be ensured that the operating voltage of
the supplied power supply unit corresponds with the existing mains voltage (100 VAC – 240
VAC).
2.
Only employ the with the MC 10
plus
supplied suitable external power supply unit, otherwise the
analyser could be damaged.
3.
It is recommended that all repairs beyond the periodical maintenance and little setting are
carried out by the ABW-customer service.
4.
If the instrument is not used as advised in the manual the safe operation is not granted and the
guarantee expires.
5.
The instrument may not be connected to an extension lead.
6.
The total length of the mains connection may not exceed 3 meters.
Warning!
Only the original external power supply (100 VAC – 240 VAC) which is
delivered with the MC 10
plus
may be used, otherwise the analyser
could be damaged.
2.6
Connection of the device
1.
Connect the low voltage cable of the power supply unit with the low voltage plug at the back of
the instrument.
2.
Insert the plug of the power supply unit into a socket.
3.
If an automatic pipette is used connect the pipette contact line to one of the according sockets
at the back of the MC 10
plus
.
4.
If an additional printer is used connect the data line of the printer with the RS-232 interface
(chapter 1.6).
or
If you require an online connection then the EDP-data line has to be connected with the RS-232
interface.
5.
The data line for the external printer or for the online connection may not be longer than 3
meters.
6.
If an external barcode scanner shall be used connect the contact line of the scanner with the
socket for the scanner (chapter 1.6).