28
D33363 | Date 06/2016
Operation and control
3. Press the confirmation key.
The display now shows the remaining time in
a large font and the approximate end time in
a smaller font beneath.
TIMER
30
m
04
h
End 13:30 23.11.
4. Now, as described under 5.4.2, set the individual values for temperature, air flap position
etc. which you want the appliance to operate at. The set values can be changed at any
time while the timer elapses. The changes are effective immediately.
In Setup, you can choose if the timer should run setpoint-dependent or not. This deter-
mines whether the timer should not start until a tolerance band around the set tempera-
ture is reached or whether it should start right after activation (see page 39). If the timer
runs setpoint-dependent, this is indicated by the
symbol in the timer display.
When the timer has elapsed, the display shows 00h:00m. All func-
tions (heating etc.) are switched off. If a fan had been active, it will
keep on running for a short safety period. In addition, an acoustic
alarm sounds, which can be turned off by pressing the confirma-
tion key.
To deactivate the timer, open the timer display by pressing the
activation key again and then turning the turn control to reduce
the timer setting until --:-- is displayed. Confirm with the confirma-
tion key.
5.5 Temperature monitoring
The appliance is equipped with a double overtemperature protection (mechanical/electronic)
in accordance with DIN 12 880. This serves to avoid damage to the chamber load and/or ap-
pliance in case of a malfunction:
Ź
Electronic temperature monitoring (TWW/TWB) (TWB only if equipped with a second tem-
//
perature sensor, option A6)
Ź
mechanical temperature limiter (TB)
TIMER
00
m
00
h
End 13:30 23.11.
TIMER
--
m
--
h
End 9:00 23.11.