12
D33363 | Date 06/2016
Setup and description
2.5 Connections and interfaces
2.5.1 Electrical connection
This appliance is intended for operation on an electrical power system with a system imped-
ance Z
max
of a maximum of 0.292 ohm at the point of transfer (service line). The operator
x
must ensure that the appliance is operated only on an electrical power system that meets
these requirements. If necessary, you can ask your local energy supply company what the
system impedance is.
Observe the country-specific regulations when connecting (e.g. in Germany
DIN VDE 0100 with residual current circuit breaker).
2.5.2 Communication interface
The Ethernet interface is intended for appliances which meet the requirements of IEC 60950-1.
Via Ethernet interface, the appliance can be
connected to a network to read out protocol
logs with AtmoCONTROL, the optional
appliance software. The Ethernet interface is
located on the rear of the appliance (Fig. 4 ).
For identification purposes, each appliance
connected must have its own unique IP ad-
dress. Setting the IP address is described on
page 36.
With an optional USB to Ethernet converter,
the appliance can be directly connected to a
computer / laptop (see Optional accessories
on page 15).
2.6
Designation ( nameplate)
The nameplate (Fig. 5) provides information about the appliance model, manufacturer and
technical data. It is attached to the front of the appliance, on the right side under the door
(see page 10).
Typ: UN 260 F.-Nr.: 0109.0088
230 V~ 14.8 A 50/60 Hz 3400 W
DIN12880-Kl.3.1 Nenntemp.: 300 °C
1
2
3
4
5
6
10
8
9
7
Fig. 5
Nameplate (example)
1 Type designation
2 Operating voltage
yp
g
3 Applied standard
p
g
g
g
g
4 Protection type
5 CE conformity
yp
6 Address of manufacturer
7 Disposal note
8 Temperature range
9 Connection / power ratings
p
g
10 Appliance number
pp
Fig. 4
Ethernet interface