D33363 | Date 06/2016  

39

Menu functions

7.3.4 Timer Mode

Here, you can choose whether the digital backwards counter with 

target time setting (see page 27, timer) should run setpoint-

dependent or not – this determines whether the timer should 

not start until a tolerance band of ±3 K around the set 

temperature is reached (Fig. 23, B) or whether it should start 

right after activation (A).

t

Timer

t

Timer

A

B

°C/°F

t

6 K

ž

ž

Fig. 23

 Timer Mode

Timer independent of setpoint: Timer starts right after activation

g

Timer setpoint-dependent: Timer does not start until tolerance band is reached

p

p

g

p

g

If the temperature tolerance band is exceeded in setpoint-dependent operation, the timer will 

be interrupted and only be resumed when the setpoint temperature has been reached again.

7.3.5  Type of the slide-in unit (Grid or Shelf)

Here, you have to set the type of the slide-in unit (grid or shelf) 

used. The selection Shelf enables you to adjust the control 

function to the different air flow characteristics in the interior 

when using optional sliding shelves instead of the grids that are 

part of the standard delivery.

7.3.6 Remote control

In the setup entry 

Remote control

, you can set whether the 

appliance should be controlled via remote control and if 

so, in which mode. These settings are available:

Ź

Off

Ź

Read Only

Ź

Write

Ź

Write+Alarm

When the appliance is in remote control mode, the 

symbol appears in the temperature display. In the 

settings 

Write

 and 

Write+Alarm

, the appliance cannot 

be controlled at the ControlCOCKPIT until the remote 

control has been switched off (setting Off) or set to 

Read

Only

.

In order to use the remote control function, program-

ming skills and special libraries are required.

SETUP

Unit

IP address

255.145.136.225

Subnet mask

255.255.0.0

Alarm temp

°C

F

TWW

TWB

Timer mode

Unit

IP adress

255.145.136.225

Subnet mask

255.255.0.0

°C

°F

Timer mode

Alarm temp

TWW

TWB

Slide-in unit

Grid         Shelf

Setup

Balance

Remote Control

   

Off

Gateway                192.168.5  .1

+30 %

2/2

23

.2°C

TEMP

Set 180

.0°C

Содержание IF 1060

Страница 1: ...UNxx UFxx INxx IFxx OPERATING INSTRUCTIONS...

Страница 2: ...mail service memmert com When contacting customer service always quote the product serial number on the nameplate see page 12 Shipping address for repairs pp g p Memmert GmbH Co KG Kundenservice Will...

Страница 3: ...r to the latest firmware version Due to individual configurations and sizes illustrations in this manual may be slightly different from the actual appearance Function and operation are identical Other...

Страница 4: ...ommunication interface e 12 2 6 Designation nameplate 12 2 7 Technical data 13 2 8 Applied directives and standards 14 2 9 Declaration of conformity y 14 2 10 Ambient conditions 15 2 11 Scope of deliv...

Страница 5: ...messages 31 6 1 Warning messages of the temperature monitoring function 31 6 2 Malfunctions operating problems and appliance errors 32 6 3 Power failure e 33 7 Menu mode 34 7 1 Overview w 34 7 2 Basic...

Страница 6: ...r somebody else could be injured if you do not observe the accompanying safety regulation Caution is used for information that is important for avoiding damage 1 1 2 Signs used Warning signs warning o...

Страница 7: ...ith materials test objects which do not form any toxic or explosive vapours when heated up see also Intended use on page 8 Warning If the door is open while the appliance is in operation the appliance...

Страница 8: ...G 24 The appliance may only be loaded with materials and substances which cannot form any toxic or explosive vapours at the set temperature and which cannot explode burst or ignite The appliance may n...

Страница 9: ...31 1 8 Switching off the appliance in an emergency Push the On Off switch on the control panel Fig 1 and disconnect power plug This disconnects the appliance from the power supply at all poles Warning...

Страница 10: ...nstruction 1 ControlCOCKPIT with capacitive function keys see page 25 2 On Off switch see page 22 y p g p g 3 Working chamber fan for p g p g UFxx IFxx appliances only g g 4 Steel grid 5 Working chamb...

Страница 11: ...loys stainless steel Mat No 1 4016 ASTM 430 and for the interior stainless steel Mat No 1 4301 ASTM 304 is used which stands out through its high stability optimal hygienic properties and corrosion re...

Страница 12: ...rotocol logs with AtmoCONTROL the optional appliance software The Ethernet interface is located on the rear of the appliance Fig 4 For identification purposes each appliance connected must have its ow...

Страница 13: ...1600 850 1100 1100 1100 1100 1500 1800 230 V 50 60 Hz 1600 1000 1250 1400 1600 1700 1800 2000 UN UF 230 V 50 60 Hz 1600 2000 2500 2800 3200 3400 115 V 50 60 Hz 1600 1700 2200 2200 2200 2200 400 V 50 6...

Страница 14: ...on of the laws of member states relating to electrical equipment designed for use within certain voltage limits Standards complied with DIN EN 61 010 1 VDE 0411 Part 1 DIN EN 61 010 2 010 VDE 0411 Par...

Страница 15: ...nce could lead to sedimentation in the interior and as a consequence could result in short circuits or dam age to electrical parts For this reason sufficient measures to prevent large clouds of dust o...

Страница 16: ...must be transported with a manual pallet jack or forklift truck 30 55 75 110 160 260 450 750 Warning You may get your hands or feet squashed when transport ing and installing the appliance Wear protec...

Страница 17: ...n edge 3 4 1 Checking for completeness and transport damage Check the delivery note to ensure that the delivery is complete Check the appliance for damage If you notice deviations from the delivery no...

Страница 18: ...he weight of the appliance see Technical data on page 13 Do not place the appliance on a flamma ble surface Depending on the model see nameplate a 230 V 115 V or 400 V power connection must be availab...

Страница 19: ...Check the load capacity first 9 9 9 9 U U U U Stacked two appliances maximum mount ing material feet provided 9 9 9 9 U U U U Wall mounting Separately packaged fastening material is included in the s...

Страница 20: ...The tilt protection is included in the delivery 1 As illustrated fasten the tilt protection to the rear side of the appliance 2 Bend the tilt protec tion upwards by 90 in the desired distance to the w...

Страница 21: ...t warp due to the floor condi tions In order to do so every door has two adjuster screws at the top and at the bottom Fig 9 First adjust the door at the top and then if further adjustment is necessary...

Страница 22: ...t does not come into contact with any hot parts 230 115 V appliances pp Plug the provided power cable into the rear of the appli ance and connect it to a CEE 7 4 socket Fig 10 400V appliances pp The p...

Страница 23: ...Opening the door To open the door pull the door handle to the side to the left or to the right depending on the door variation see Fig 13 A and open the door completely To close the appliance push th...

Страница 24: ...are specified in the technical data overview on page 13 The chamber must not be loaded too tightly so that proper air circulation in the working chamber is guaran teed Do not place any chamber load on...

Страница 25: ...r temperature setpoint adjustment 2 Setpoint and actual temperature display j 3 Fan speed display p 4 Activation key for fan speed setting p p y p 5 Switch to menu mode see page 34 y p g y 6 Activatio...

Страница 26: ...ordingly If no new values are entered or confirmed for approx 30 seconds the appliance automati cally returns to the main menu and restores the former values If you want to cancel the setting procedur...

Страница 27: ...ed fresh air operation in steps of 10 FLAP 40 Fan speed p only for UFxx IFxx appliances Adjustment range 0 to 100 in steps of 10 FAN 50 5 4 5 Operation with digital backwards counter with target time...

Страница 28: ...is indicated by the symbol in the timer display When the timer has elapsed the display shows 00h 00m All func tions heating etc are switched off If a fan had been active it will keep on running for a...

Страница 29: ...over temperature control and begins to regulate the monitoring temperature TWW Fig 17 or switches off the heating TWB Fig 18 The two types of temperature monitoring are only available for appliances...

Страница 30: ...arm is additionally signalled by an intermittent acoustic signal which can be turned off by pressing the confirma tion key Information on what to do in this case are provided in the chapter Malfunctio...

Страница 31: ...ller TWW has assumed heat ing control Increase the difference between the monitoring and setpoint temperature by either in creasing the max value of the temperature monitoring MAX temp or decreasing t...

Страница 32: ...f the timer or switch off the remote control Displays suddenly look different Appliance is in wrong mode Change to operating or menu mode by pressing the MENU key Error message E 3 in the temperature...

Страница 33: ...HS TMSHK SGD OOKH MBD BNNKR CNVM QRS In case of a power failure the appliance operates as follows In manual mode After power supply has been restored operation is continued with the parameters set The...

Страница 34: ...accepted by pressing the confirmation key are saved 7 1 Overview Press the MENU key to change between the displays in menu mode SETUP SIGNALT NE CALIB LANGUAGE TIME 1 2 3 4 5 6 7 8 9 10 11 Fig 21 Cont...

Страница 35: ...gain press the activation key which you have used to activate the display The appliance returns to the menu overview Only the settings that you have con firmed by pressing the confirmation key before...

Страница 36: ...slide in unit Grid or Shelf see page 39 Remote control see page 39 Gateway see page 40 If the SETUP menu contains more entries than can be displayed this is indicated by the display 1 2 This means th...

Страница 37: ...SETUP Unit IP address 255 168 100 100 Subnet mask 255 255 0 0 Alarm temp C F TWW TWB Timer mode 5 After setting the last three digits accept the new IP address by pressing the con firmation key The se...

Страница 38: ...Subnet mask 255 255 0 0 Alarm temp C F TWW TWB Timer mode Slid i it 4 Save the setting by pressing the confir mation key SETUP Unit IP address 255 145 136 225 Subnet mask 255 255 0 0 Alarm temp C F T...

Страница 39: ...lection Shelf enables you to adjust the control function to the different air flow characteristics in the interior when using optional sliding shelves instead of the grids that are part of the standar...

Страница 40: ...he right side of the Time display The display is enlarged and the first adjustment option Date automatically highlighted Date and time Time Time zone Daylight savings 12 05 2012 12 00 GMT 01 00 Date 2...

Страница 41: ...s case on 9 Save the setting by pressing the confirmation key Time 12 05 2012 12 00 GMT 00 00 Date Time zone Daylight savings Daylight saving time and standard time are not changed automatically For t...

Страница 42: ...ed reference measuring device CAL 1 0 5 K CAL 3 0 8 K CAL 2 2 6 K 180 C 40 C 120 C 0 C factory calibration Fig 24 Schematic example of temperature adjustment Example Temperature deviation at 120 C sho...

Страница 43: ...and in manual mode adjust the set temperature to 120 C 21 4 C TEMP 120 0 C Set 8 Wait until the appliance reaches the set temperature and displays 120 C The reference instrument for example displays...

Страница 44: ...teel housing Rust deposits can lead to an infection of the stainless steel If rust spots should appear on the surface of the working chamber due to impurities the affected area must be immediately cle...

Страница 45: ...be disposed of in normal household waste For disposal please contact your dealer or the manufacturer Any appliances that are infected infectious or contami nated with materials hazardous to health ar...

Страница 46: ...ency 9 Ending operation 30 Error description 32 Error messages 31 Explosion protection 8 F factory calibration 43 Fan speed 27 Fehlermeldung 32 Forklift truck 17 Function 11 G Gateway 40 Graph 30 Grid...

Страница 47: ...re 27 Temperature comparison 42 Temperature deviation 42 Temperature limiter 30 Temperature monitoring 28 37 Time 39 Timer 27 Timer mode 39 Transport 16 17 Transport damage 17 Turn control 26 TWB 28 T...

Страница 48: ...t GmbH Co KG Willi Memmert Stra e 90 96 D 91186 B chenbach Tel 49 9122 925 0 Fax 49 9122 14585 E Mail sales memmert com facebook com memmert family Die Experten Plattform www atmosafe net Universal ov...

Отзывы: