42
D33363 | Date 06/2016
Menu functions
7.5 Calibration
The appliances are temperature calibrated and adjusted at the factory. In case readjustment
should be necessary later on – for example due to influence of the chamber load – the ap-
pliance can be calibrated customer-specifically using three calibration temperatures of your
choice:
Ź
Cal1
Temperature calibration at low temperature
Ź
Cal2 Temperature
calibration at medium temperature
Ź
Cal3
Temperature calibration at high temperature
To guarantee perfect control, we recommend to calibrate the appliance once a year.
For temperature adjustment, you will need a calibrated reference measuring device.
CAL 1
+0,5 K
CAL 3
-0,8 K
CAL 2
+2,6 K
180°C
40°C
120°C
0°C
factory calibration
Fig. 24
Schematic example of temperature adjustment
Example: Temperature deviation at 120 °C should be corrected.
1. Activate the adjustment setting. To do so,
press the activation key on the right side
of the
CALIB
display. The display is en-
larged and the first calibration tempera-
ture – in this case 40 °C – automatically
highlighted.
Cal1
°C
-0,2
K
Cal2
100.0
°C
+0,1
K
Cal3
180.0
°C
-0,2
K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Calibration
Temperature
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
100.0
180.0
40.0
2. Press the confirmation key repeatedly,
until the calibration temperature Cal2 is
selected.
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Temperature
Cal1
Cal2
Cal3
180.0
40.0
100.0