background image

HCP

page 30

40°C

34°C

ASF alarm

ASF activated

ASF activated

40°C

34°C

ASF-alarm

ASF activated

ASF activated

33°C

15°C

°C

27°C

ASF set to

+/-3°C in

SETUP

13 Active humidity control (%rh)

13 Active humidity control (%rh)

Active humidity control ensures that the set humidity is reached rapidly without the use ofwater dishes.
During the heating-up phase the humidity control is initially de-activated. Humidification and 
dehumidification control starts to operate approx. 5 minutes after the set temperature has been reached. 
The setpoint can be adjusted between 20% rh and 95% rh. The humidity setpoint can be set or altered 
during the stabilisation phase. Humidification operates by passing water vapour through a dosing pump 
into the working chamber. In order to prevent bacteria formation the steam is first heated to approx. 
140°C. De-humidification consists of the supply of dry air through sterile filter.
If no humidity is required in the chamber, humidity control can be de-activated by the setting 

OFF

.

is indicated if the humidity exceeds the selected setpoint for longer than 
30 minutes. The humidity indication then flashes as well as the symbol 
. In this case open the door for 30 sec and wait whether the controller 
controls in a stable manner on the setpoint. If the fault is repeated, contact 
customer service.
(Note: this monitor function becomes activated only after the humidity 
setpoint has been reached for the first time)
is indicated when the water container is empty.

Humidification is interrupted automatically when the outer door is opened. 
The text display shows 

DOOR open

.

Setpoint changed = ASF de-activated automatically

Setpoint changed = ASF de-activated automatically
If the temperature setpoint is altered, the automatic temperature monitor is automatically de-activated 
temporarily (see in the example the setpoint is changed from 37°C to 30°C) until the tolerance margin of 
the new temperature setpoint is reached (see in the example below: the ASF is re-activated at 30°C, +/- 
3°C).

Содержание HCP 108

Страница 1: ...OPERATING INSTRUCTIONS Humidity Chambers HCP 108 HCP 153 HCP 246...

Страница 2: ...HCP page 2...

Страница 3: ...ing mode 13 6 Setting the parameters 13 7 Normal operation 14 8 Weekly programmer 16 9 Programme operation 18 9 1 Closure commands for ramp segments 20 10 Printer 24 11 Basic oven settings Setup 25 11...

Страница 4: ...9 Report memory 40 19 1 Reading the report memory 40 19 2 Reading the report memory into the PC via RS232C 40 19 3 Printing the report memory from the oven 40 20 Cleaning 41 21 Maintenance 42 22 Check...

Страница 5: ...arantee and indemnification are excluded This mark on the product means Note Operating Instructions Note Operating Instructions Warning oven hot when operating Warning oven hot when operating The righ...

Страница 6: ...the oven 2 1 Subframe accessory 2 1 Subframe accessory Models 108 153 and 246 can be placed on a subframe 2 2 Stackable version accessory 2 2 Stackable version accessory Two ovens of the same model si...

Страница 7: ...Specification VBG 24 and are therefore not suitable for drying evaporating and burning in of paints enamels or similar materials whose solvents may produce an inflammable mixture with air There must...

Страница 8: ...s measued with Pt100 in 4 wire circuit adjustment range normal operation with humidity 20 C to 90 C normal operation without humidity 20 C to 160 C sterilisation operation 160 C 4 hours via STERICard...

Страница 9: ...ses at the coldest point of the unit due to the dew point With prolonged operation outside the working range there is a With prolonged operation outside the working range there is a possibility of wat...

Страница 10: ...fan disc Integral weekly programmer with group function e g all workdays Two separate PT100 temperature sensors Class A in 4 wire circuit for control and monitoring Digital monitor controller for over...

Страница 11: ...oduct is intended to operate on a supply network with a system impedance Zmax at the transfer point max at the transfer point max building connection of 0 292 Ohm max The user has to ensure that the p...

Страница 12: ...e turbulence free chamber ventilation 3 1 1 1 1 1 2 3 4 1 Operating the door 4 1 Operating the door The door is opened by clockwise rotation of the handle Close the door by turning the door handle ant...

Страница 13: ...the controller operates in the new operating mode PRINT SETUP Normal operation Weekly programmer Programme operation Printer Basic settings 6 Setting the parameters 6 Setting the parameters After an...

Страница 14: ...sh turn control the following parameters can be selected and can be altered as described in the section 6 Temperature setpoint Temperature setpoint Adjustment range with humidity 20 to 90 C without hu...

Страница 15: ...ntil the overtemperature protection and the symbol MAX or MIN flashes Keep the SET key depressed and use the push turn control to set the overtemperature protection to 51 0 C and the undertemperature...

Страница 16: ...t switched on without an ON time being entered an OFF time cannot be programmed for the respective time block either on off ON time Adjustment range 00 00 to 23 59 hrs on off OFF time Adjustment range...

Страница 17: ...ch on at 9 30 hrs Mo Fr Turn the push turn control anticlockwise to select Mo Fr on workday group Hold down the SET key and use the push turn control to set the switch on time to 9 30 on off Mo Tu We...

Страница 18: ...n the Section 6 loop t3 t4 t2 t1 on off Mo Tu We Th Fr Sa Su 3 4 2 1 STERI DEFRO mb IN 1 IN 2 OUT MIN AUTO MAX C C rh h PRINT SETUP Delayed programme start switch on day Delayed programme start switch...

Страница 19: ...FRO mb IN 1 IN 2 OUT MIN AUTO MAX C C rh h PRINT SETUP Humidity setpoint humidity at end of the ramp segment Humidity setpoint humidity at end of the ramp segment Range 20 to 95 rh loop t3 t4 t2 t1 on...

Страница 20: ...next programme segment only when the programmed temperature setpoint and the programmed humidity setpoint have been reached even if the programmed heating time has already elapsed LOOP Ramp repeat fun...

Страница 21: ...e push turn control After the SET key has been released the controller is in operating mode programme operation PRINT SETUP 2 Edit programme 2 Edit programme Hold down the SET key and turn the push tu...

Страница 22: ...ication Hold down the SET key and set the time 00 45 using the push turn control h 10 Select temperature of second ramp segment 10 Select temperature of second ramp segment Turn the push turn control...

Страница 23: ...rh 16 Set closure command for third ramp segment 16 Set closure command for third ramp segment Turn the push turn control clockwise until a segment closure command e g end appears Press the SET key b...

Страница 24: ...section 19 The report data can in this operating mode be printed out through the printer connected to the oven When using a colour printer the various graphics can be printed in colour On the printou...

Страница 25: ...matically but must be set manually by the user Date The controller incorporates a calendar which automatically allows for the different lengths of the months and also for leap years Weekday Weekday Tu...

Страница 26: ...ature protection mechanical electronic according to DIN 12880 12 1 Mechanical temperature monitor temperature limiter TB 12 1 Mechanical temperature monitor temperature limiter TB All ovens of the HCP...

Страница 27: ...N AUTO MAX Using the push turn control select the symbol MIN Hold down the SET key and set the protection temperature using the push turn control Note The temperature monitor can be set independently...

Страница 28: ...n control Automatic temperature monitor ON ASF ON C MIN AUTO MAX Using the push turn control select the symbol AUTO Hold down the SET key and select on using the push turn control Notes on the ASF The...

Страница 29: ...ple through opening the oven door during operation triggers the alarm Triggering the ASF alarm is indicated by flashing symbol AUTO and If the sounder is switched on in SETUP the ASF alarm is addition...

Страница 30: ...ilter If no humidity is required in the chamber humidity control can be de activated by the setting OFF is indicated if the humidity exceeds the selected setpoint for longer than 30 minutes The humidi...

Страница 31: ...activated if the overtemperature protection operates see section 12 2 3 lo alarm undertemperature alarm activated if the undertemperature protection operates see section 12 2 3 asf alarm temperature o...

Страница 32: ...e automatic sterilisation programme can not be altered by the user After successful sterilisation the control displays STERILISATION OK For documentation purposes the sterilisation process can be repo...

Страница 33: ...orresponding serial number Each log on via the User ID Card is documented in the internal flash memory If the USER IDcard is inserted the SETUP menu includes the additional item ID LOCK When the setti...

Страница 34: ...the measured reference temperature is too low the calibration correction setting has to have a negative sign 4 Carry out a check measurement using the reference meter 5 The procedure can be carried o...

Страница 35: ...e factory temperature calibration is restored C C C C C C calibration point1 calibration point 3 calibration point 2 calibration temperature calibration temperature calibration temperature calibration...

Страница 36: ...in SETUP If the measured reference humidity is too low the calibration correction setting has to have a negative sign 4 Carry out a check measurement using the reference instrument 5 The procedure ca...

Страница 37: ...op t3 t4 t2 t1 on off Mo Tu We Th Fr Sa Su 3 4 2 1 STERI DEFRO C C rh mb CO mb 2 IN 1 IN 2 OUT IN 1 IN 2 OUT MIN AUTO MAX PRINT SETUP loop t3 t4 t2 t1 on off Mo Tu We Th Fr Sa Su 3 4 2 1 STERI DEFRO C...

Страница 38: ...h the oven The default setting is ADDRESS 0 Using this address each oven can be addressed by the PC and programmed If several ovens are connected by the RS232C interface to one PC each oven requires a...

Страница 39: ...communicates with the oven The default setting is ADDRess 0 Using this address each oven can be addressed by the PC and programmed For this purpose the PC must be equipped with an RS485 interface or m...

Страница 40: ...nted either via the RS232C interface or by a PLC3 compatible printer connected to the oven 19 2 Reading the report memory into the PC via RS232C 19 2 Reading the report memory into the PC via RS232C U...

Страница 41: ...ternal casing If rust stains caused by contamination occur on the interior surface the affected areas must be cleaned and polished immediately ATTENTION Do not use cleaning fluids which contain scouri...

Страница 42: ...osure of the door is ensured by a seal on the oven and another one on the door In continuous operation the flexible sealing material may take a permanent set Readjustment may then be necessary in orde...

Страница 43: ...SF has been activated and rh empty water container empty and rh over humidity setpoint has been exceeded E 0 in temperature module error on self test E 1 in temperature module power unit triac faulty...

Страница 44: ...After a supply failure lasting less than 60 minutes the current programme is continued at the point where it was interrupted The instant and duration of the supply failure are documented in the repor...

Страница 45: ...61326 2004 05 EN 61326 1997 EN 61326 A1 1998 EN 61326 A2 2001 EN 61326 A2 2003 The designated product is in conformity with the European Low Voltage Directive 2006 95 EEC including amendments Council...

Страница 46: ...f domestic refuse For disposal please contact your dealer or the manufacturer Products which are infected infectious or contaminated with health endangering substances are excluded from return Please...

Страница 47: ...page 47...

Страница 48: ...Memmert GmbH Co KG Postfach 1720 D 91107 Schwabach Tel 49 0 9122 925 0 Fax 49 0 9122 145 85 E Mail sales memmert com www memmert com 30 09 2017 HCP englisch D10345...

Отзывы: