17.2 Humidity
17.2 Humidity
User-calibration of the unit at the controller, using two humidity points:
RH 20
Humidity calibration at 20% relative humidity
RH 90
Humidity calibration at 90% relative humidity
Either a positive or a negative calibration correction can be applied to each selected calibration point.
General calibration instructions:
1.
Select the required calibration humidity in SETUP and set the corresponding calibration
correction to
0.0
°C.
2.
Measure the deviation from the selected calibration humidity under steady conditions, using a
reference instrument.
3.
Set the calibration correction in SETUP. If the measured reference humidity is too low, the
calibration correction setting has to have a negative sign.
4.
Carry out a check measurement using the reference instrument.
5.
The procedure can be carried out for up to 2 calibration humidities.
Example: Humidity deviation at the load at 90% has to be corrected:
1.
Set calibration humidity in SETUP on
RH 90
and set the corresponding calibration correction
to
0.0
%
2.
Using a calibrated reference instrument, an actual humidity of 88% is measured in normal
operation for a humidity setting of
90
%.
3.
In SETUP set calibration correction for
RH 90
to
-2.0
%
4.
The reference instrument should read 90.0% after the unit has settled down.
5.
An additional calibration can be programmed with
RH 20
at 20% relative humidity.
rh 20
+3
%
rh 90
-2 %
factory
calibration
60%
40%
0%
20%
80%
100%