53
Error message
Possible causes
What you can do
has been started in overpressure, an additional
request appears to sterilize the sterile filter,
since this has become damp and is possibly
contaminated with microbes. Exchange sterile
filter at the back side of the steam sterilizer,
sterilize the filter in the
クイック
フ゜ロク゛ラ
ム
,
after the expiration of the program, plug in
the filter again
Disconnection of the steam
sterilizer during a running
program
Only terminate a running program with K
EY
(S)
(see also page 25,
Manually terminate
program
)
エラー
33
:
アツリョク コウ
カ
The maximum switch-on time of
the steam generator for
achieving the control pressure
was exceeded
On repeated occurrence notify your specialist
dealer
エラー
34
:
メッキン
TU1
Shortfall of the minimum
permissible sterilization
temperature
(temperature sensor 1)
Operate device with smaller load, possibly
carry out vacuum test
Check door seal for wear
On repeated occurrence notify your specialist
dealer
エラー
35
:
メッキン
TO1
Exceeding the maximum
permissible sterilization
temperature
(temperature sensor 1)
Carry out vacuum test
On repeated occurrence notify your specialist
dealer
エラー
36
:
メッキン
PU
Shortfall of the minimum
sterilization pressure
Operate device with smaller load, possibly
carry out vacuum test
Check door seal for wear
On repeated occurrence notify your specialist
dealer
エラー
37
:
メッキン
PO
Exceeding the maximum
permissible sterilization
pressure
On repeated occurrence notify your specialist
dealer
エラー
38
:
メッキン
TD1
Maximum permitted difference
between theoretical
temperature, quoted from the
pressure signal, and the
temperature measured at the
temperature senor 1 was
exceeded
On repeated occurrence notify your specialist
dealer
エラー
41
:
ソウフウ カンソ
ウ
Flow filter soiled
Check flow filter lower front on the chamber
(right behind the door)
On repeated occurrence notify your specialist
dealer
エラー
42
:
カンソウ カアツ
ホ゜ンフ゜
Sterile filter soiled
Check sterile filter, exchange if necessary
On repeated occurrence notify your specialist