Chapter 3
– Initial start-up
15
Space requirements
Euroklav
29 VS+
Width
Height
Depth
A= 42.5 cm
B= 48.5 cm
C= 61 cm
D= 32 cm
E= 12 cm
F= 10 cm
G= 10 cm
maximum acoustic power for Euroklav
29 VS+:
65 dB (A)
Additional space for
feed water supply
Other than for the steam sterilizer, the water treatment unit for
the feed water supply require space.
NOTICE
In no case may you subject the water storage tank to sunlight.
This way you avoid algae formation.
The maximum suction height from the external storage tank is
1.5 metres.
Space requirement for water treatment unit
MELA
dem
40
can be directly hung on to the steam sterilizer on the side
The attachment can be next to the steam sterilizer or
above or below the steam sterilizer
MELA
dem
47
Attachment next to the steam sterilizer or above or below
the steam sterilizer (e.g. lower cabinet)
delivered with water storage tank
Water treatment unit
Width
Height
Depth
Diameter
MELA
dem
40
32 cm
35 cm
15 cm
-
MELA
dem
47 module housing
39 cm
47 cm
15 cm
-
MELA
dem
47 water storage tank
-
51 cm
-
24 cm
Connections required
Danger of injury!
■
The electric connections and connections for feed water
and used water may only be completed by a specialist
technician.
Non-observance could lead to a short-circuit and/or fire
and/or water damage and/or electrical shock. Serious
injuries could be the consequence.