TM1700-sEriEs
ZP-BL16E
BL1448GE
4 CABA LOCAL
Defining pressure transducers
1]
Press the “Pressure” button on the “DEFINE
NEW TRANSDUCER” screen.
Calibrating pressure transducers can be done in two
different ways:
A]
Manual. Enter all values (pairs of pressure &
voltage or current value) manually.
B]
Using the instrument readings as reference.
Enter pressure value manually while using au-
tomatic TM 1800 voltage or current reading
A. Manually calibrating a pressure
transducer
1]
Select the appropriate transducer output
type and if needed the power supply by
pressing the “Output/Supply” button repeat-
edly.
2]
Press the “Calibrate” button. The following
question appears.
3]
Press the “NO” button. The following screen
appears.
4]
Enter the calibration values in the fields
above the graphic. For quick navigation be-
tween the fields use the arrow buttons next
to it.
5]
Type in a significant name for the transducer
into the “Transducer ID” field.
6]
Press the “Finish Calibration” button.
7]
Press the “BACK” button to leave this screen.
B. TM1700 as reference for calibrating a
pressure transducer
1]
Press the “Calibrate” button. The following
question appears.
2]
Press the “YES” button. The following screen
appears.
3]
Press the “CANCEL” button if those settings
need to be corrected and start at point 1
again. Press the “OK” button if the settings
are correct. The following message appears.
Содержание TM1700 Series
Страница 2: ......
Страница 15: ...BL1448GE ZP BL16E TM1700 series 15 3 System overview...
Страница 25: ...BL1448GE ZP BL16E TM1700 series 25 4 CABA Local...
Страница 61: ...BL1448GE ZP BL16E TM1700 series 61 5 Perform testing...
Страница 75: ...BL1448GE ZP BL16E TM1700 series 75 7 Specifications...
Страница 85: ...BL1448GE ZP BL16E TM1700 series 85 10 Quick Guide...
Страница 91: ......