Unit Maintenance
MPQ1000-UG-EN Rev 6 April 2020
85
8.
Place the battery in the battery compartment with the wires under it.
9.
Put the battery compartment door back in place and tighten the retaining screws.
10.
Put the stand back in place.
11.
Turn on the MPQ1000 by pressing and releasing the ON/OFF button. The battery
status indication should indicate 0%. The battery status indicator must be initialized
before it will display the true charge level of the battery. This is done by discharging
the battery to below 10% of its capacity.
12.
Go to the SETUP > PREFERENCES screen by pressing the HOME key, scrolling
down to the SETUP menu item and pressing the ENTER
key and then scrolling
down to the PREFERENCE menu item and pressing the ENTER
key.
13.
Select the AUTO-OFF box and, if necessary, toggle it to its un-checked state by
pressing the ENTER
key. This will allow the unit to remain on even when not
being actively used.
14.
Allow the unit to run until the battery status indicator shows a value of between 1
and 9%. The unit may be used during this discharge period, if desired; but, it should
be noted that the battery charge level indicated on the display will not be accurate
until the battery level has fallen below 10% capacity. If desired, the unit can be left
running and allowed to fully discharged and shut down.
Battery
Compartment
Floor
Battery
Compartment Wall
Battery
Connector