MPQ1000-UG-EN Rev 6 April 2020
42
Battery
Battery Charging
Always ensure that the battery is properly charged. It must maintain a 12-volt minimum
for proper recording operation. A battery charge indicator on the MPQ1000 display
allows you to view the batteries state of charge.
NOTE:
The battery should be charged prior to recording with the unit powered off.
Charging the MPQ1000 Battery
A DC adapter is provided as standard equipment with the MPQ1000. This adapter both
powers the unit and charges the internal NiMH batteries within the unit.
Plug the AC adapter into the MPQ1000 unit using the DC input jack. Then plug the
adapter into a standard AC outlet (110V-240VAC 50Hz/60Hz). The battery should fully
charge in approximately 2 hours.
Battery Storage
The MPQ1000 NiMH battery self-discharges an average of 1%/ day at room
temperature. If left for 90 days at room temperature, the cells can fully discharge. Cell
reversal can occur, which will damage the cells.
Fully charge the battery before storage. Store it in a cool dry area. Recharge battery pack
within 90 days after initial storage - sooner if stored above 30C.
Installation
1.
The MPQ1000 is an IP54 rated device. Ensure that the MPQ1000 PQ Analyzer is
not exposed to water, excessive dust, and other hazardous conditions. The
MPQ1000 PQ Analyzer is not meant for outdoor use.
The MPQ1000 comes with an optional strap for hanging the analyzer.
2.
Plug the AC adapter into the MPQ1000 unit using the DC input jack. Then
plug the adapter into a standard AC outlet (110V-240VAC 50Hz/60Hz).