DET2/3
15
Earth / ground resistance
Manual mode
Continuous graphical mode
4
G
F
V
I
C1
E
E/X
CA
T IV 300V
CA
T IV 300V
CA
T IV 300V
CA
T IV 300V
Y
V
P1
ES
P2
S
C2
H
61.8% of EC
38.2% of EC
Electrode under test
Potential spike
Current spike
P1
P2
C2
G
F
V
I
C1
E
E/X
CA
T IV 300V
CA
T IV 300V
CA
T IV 300V
CA
T IV 300V
Y
V
P1
ES
P2
S
C2
H
61.8% of EC
38.2% of EC
Electrode under test
Potential spike
Current spike
P1
P2
C2
Connect test leads to instrument
Connect test leads to instrument
5
Set test parameters
Set test parameters
Note:
15/50V
15 or 50 test voltage
Manually select a frequency
Hz
Automatically select a frequency
Noise filter on or off
6
Start test
Start test
7
Press
TEST
Press and hold
TEST
until padlock icon appears
Press
TEST
: Stop the current data stream and restart the graph.
: Activate / deactivate the average display
: Go back to test settings (the Ready screen)
8
End Test
End Test
9
Test ends
Press to stop
TEST
Press to stop
TEST
OR
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30