DET2/3
11
Operation
Test leads and terminal connections
Test lead set-up and terminal connections are described as part of the test procedure.
Important:
When the instrument is connected to electrodes, make sure that all leads and cables are fully
unwound and laid out without loops.
Important:
When running test leads out to each remote spike, take care not to lay them close to each other.
This is to minimise the effect of mutual inductance. Test leads must be spread out at least one metre apart.
Test modes
The instrument can operate in two modes to do a test:
Manual mode
In manual mode the test result can be shown as a single digital read out result, or as a continuously updated
digital read out result.
Continuous graphical mode
In continuous mode a continuous updated graph is displayed.
Green
line: Measurement line
Black
Line: Recorded average