User Interface
1006en01
03-11-24
65
CA620-CellGuard
1.
Enter PROG 1 to set normal ranges for “Blood mode” as shown above.
or
2.
Enter PROG 2 to set normal ranges for “PLT concentrate mode”.
The following example sets the normal range limits for “PLT concentrate mode”.
Note
that several parameters do not have normal ranges set as they are blocked from
being displayed and printed. See also menu 5.10.12 'Block parameters'.
Enter the required ranges by using the numerical key pad.
3.
Validate with [Enter].
4.
Scroll through this menu using the arrow keys.
5.
Press [Menu] to return.
Set floating discr. PLT/RBC
Scroll with the arrow keys to menu 5.6 and press [Enter].
In this menu the floating range for the discriminator (also called threshold) is set
for each PROG. It is not advisable to alter the PROGs default settings. The de-
fault is set to values that are strongly recommended.
Blood mode
In the example below, PROG 1 is shown by pressing [Enter] and the settings for
the floating discriminator are displayed.
The displayed fields are:
a)
PLT-L. The lowest possible setting for the floating threshold between PLT
and RBC in fl (= µm
3
). As default this is set to 15 fl.
b)
PLT-H. The highest possible setting for the floating threshold between PLT
and RBC in fl (=µm
3
). As default this value is set to 30 fl, which is the max-
imum level available in this menu.
Note:
Applicable for CA620-CellGuard
1083en.gif
1084en.gif
1086en.gif
1085en.gif
Содержание CA530
Страница 2: ......
Страница 6: ...6 03 11 24 ...
Страница 24: ...Specifications 24 03 11 24 1001en01 ...
Страница 38: ...Parameter Flags 38 03 11 24 1002en01 ...
Страница 80: ...Warning Displays 80 03 11 24 1008en01 ...
Страница 92: ...QC and Blood Controls 92 03 11 24 1156en01 ...
Страница 98: ...Maintenance Shut Down Transport 98 03 11 24 1011en01 ...
Страница 107: ......