Trouble Shooting
1012en01
03-11-24
101
3.
In case the above does not help to reduce the PLT background to accept-
able limits, contact your authorized distributor and request bulletin 000-80-
020 where instructions are given how to decontaminate the instrument.
14.7
Indication Number XXX Displayed
In case an indication number is displayed, the sample memory can still be read.
Any printout can be performed as well.
Note: An indication number is cancelled by pressing the [CE] key.
Detailed instructions regarding Indication numbers are given in the service man-
ual at your authorized distributor.
Note: The instrument will stop immediately.
Further sample processing is disabled until the instrument is disconnected and re-
connected to the mains outlet and the error is cancelled as described below.
Indication numbers in the instrument can be caused by:
1.
A mains power failure during any instrument cycle (count, prime etc.).
2.
Turning valve motor failure.
3.
Syringe motor failure.
4.
Waste pump capacity too low.
5.
Blockage due to large particles in the mixing cup device.
The following will give the user a first-aid in case an Indication number is dis-
played:
Indication = 1
Date and Time not set. Set within the SETUP Menu.
Indication = 2
The Cap Piercing Device was started and immediately aborted by turning the
safety cover Counter-Clock-Wise.
Press [CE] and the instrument will be operational.
Indication = 9
Auto HGB adjustment failed. The cause might be a broken lamp. Contact service
department.
Indication = 100-199
One of the internal motors or motor drivers failed. Restart the instrument, cancel
any indication with [CE]. If error remains, contact service department. In case of
a wrong installation (no CVT in use) this indication might occur due to excessive
variation of the mains supply. See
Mains Supply Environment
on page 39.
Indication = 200-299
The diluent status in the internal mixing cup was incorrect. Restart the instru-
ment, cancel any indication with [CE] and do a “Prime” from menu 8.1
These errors might be displayed in case there was an Indication 300-399 previ-
ously.
Indication = 300 - 399
There was a power failure in a cycle, or a cycle was aborted due to a previous in-
dication.
Содержание CA530
Страница 2: ......
Страница 6: ...6 03 11 24 ...
Страница 24: ...Specifications 24 03 11 24 1001en01 ...
Страница 38: ...Parameter Flags 38 03 11 24 1002en01 ...
Страница 80: ...Warning Displays 80 03 11 24 1008en01 ...
Страница 92: ...QC and Blood Controls 92 03 11 24 1156en01 ...
Страница 98: ...Maintenance Shut Down Transport 98 03 11 24 1011en01 ...
Страница 107: ......