5 Installation
5.1
Unpacking the Instrument
The instrument is packed as standard in a especially designed protective box.
Before the box is opened, check for any physical damages on the outside and no-
tify your carrier immediately in case of such.
Unpack the instrument and check that the following items are included:
List of materials included in delivery
Users Manual
1
Waste tube
1
Inlet tube with detector for Isotonic diluent.
1
Inlet tube with detector for hemolyzing reagent.
1
Mains cable
1
Check that the instrument and the accessories are not physically damaged.
If there is any damage or accessories are missing, contact distributor and
carrier immediately.
Installation checklist (to be sent back to the return ).
List of materials not included in standard delivery
Printer and printer- or computer connection cable.
Printer paper.
Additional software for patient sample managing systems.
Barcode reader.
Reagents, blood controls, calibrators and cleaners.
5.2
Working Conditions
Mains Supply Environment
The instrument should be operated at an indoor location only. The instrument is
designed to be safe for transient voltage as defined in IEC 801-4.
In case higher transient voltage, or mains voltage that 15% of the mark-
ing at the serial number plate are expected (e.g. within tropical areas), a CVT
(Constant Voltage Transformer, also called “Magnetic Stabilizer” or “Ferro-Res-
onant Transformer”) must be installed to protect the instrument against damage.
Guidelines are given in the “Service Manual” section “Installation auxilliary de-
vices”. Contact your authorized distributor in such a case.
An abrupt interruption of the mains supply will not damage the instrument as all
calibration constants and other parameters necessary for the correct working of
the instrument are protected against mains supply loss.
Important
The following procedures
must be followed exactly.
Boule has no responsibility
in case of faulty or errone-
ous installation. Possible
errors that may occur are:
- Indication numbers
- Erroneus parameter re-
sults
- Excessive service needs
Warning
Electrical shock hazard
Installation of external
electrical equipment such
as CVT must only be car-
ried out by authorized ser-
vice engineers. Violating
this might result in injuries
and/or loss of life and/or
erroneous parameter re-
sults.
Warning
Electrical shock hazard
The instrument must only
be connected to a ground-
ed mains supply. Violating
this, might result in injuries
and/or loss of life and/or
erroneous parameter re-
sults.
1003en01
Содержание CA530
Страница 2: ......
Страница 6: ...6 03 11 24 ...
Страница 24: ...Specifications 24 03 11 24 1001en01 ...
Страница 38: ...Parameter Flags 38 03 11 24 1002en01 ...
Страница 80: ...Warning Displays 80 03 11 24 1008en01 ...
Страница 92: ...QC and Blood Controls 92 03 11 24 1156en01 ...
Страница 98: ...Maintenance Shut Down Transport 98 03 11 24 1011en01 ...
Страница 107: ......