Installation
1003en01
03-11-24
43
3.
Locate the reagent inlets at the rear of the instrument, see picture below.
4.
Connect the corresponding tubes from the reagent probes and the level de-
tector cables as shown in the picture below.
5.
Connect the waste outlet to a proper container or open drain.
See
Waste connection
on page 40.
Final set-up
1.
Locate the serial number plate at the rear and check that the mains voltage
and frequency corresponds to your local mains supply outlet.
2.
Insert the power cable and connect to the mains supply. A short beep is
heard where after the instrument will perform a self-check that normally
takes 2 minutes. (Maximum15 minutes in case of condensation).
1028.tif
1029.tif
Caution
Bio hazard
The end of the waste tube
must be at a lower level
than the instrument itself.
Not following this might
lead to improper instru-
ment functions and/or
waste liquid flowing back-
wards into the instrument.