Routine Operation
58
03-11-24
1005en01
6.
Insert the MPA into its holder and the instrument will automatically start
the analyzing sequence.
Note:
7.
Detailed movie shots are available on Boule’s support server how to use the
MPA system in detail. Please put your browser to the following address and
download the MPEG movie shots. The movie shot shows also how to clean
the Micro Pipette before it is loaded into the MPA.
www.medonic.se/MPA/
7.5
Analysing the Sample
(Cap Piercing Device)
The cap piercing device model 220 is an optional built-in unit that enables the in-
strument to be used with several types of closed tubes. A tube selector is used
with 6 positions giving the operator the ability to use 6 different kinds of tubes.
A motor driven needle is penetrating the closed tube and aspirating the sample
after which the needle and aspiration tube is cleaned and dried.
The closed tube might be of a vacuum or zero pressure type. The adapter handles
even a slight over pressure in the closed tube in a correct way. The minimum vol-
ume in the closed tube should be ca. 1 ml using a standard B&D tube with a
length of ca. 73 mm.
1.
Rotate the wheel to a position that fits the dimension of the tubes you are
using. The positions can hold auxiliary adapter devices that can be ordered
from Boule to fit your local needs.
1065.jpg
Important
Fill the micro pipette com-
pletely and wipe off the ex-
cessive blood on the
outside surface.
Ignoring this instruction
might cause incorrect and
non-reproducable results.
Important
Maximum length of tube
77 mm. Lengths exceeding
this dimension will not fit
into the adapter.
Use only tubes with a max-
imum length of 77 mm
Warning
Contamination hazard if
contaminated blood enters
into open cut.
Always wear protective
gloves when handling
blood samples
(Gloves in these pictures
are not shown for clarity
reasons only.)
1066.bmp
Содержание CA530
Страница 2: ......
Страница 6: ...6 03 11 24 ...
Страница 24: ...Specifications 24 03 11 24 1001en01 ...
Страница 38: ...Parameter Flags 38 03 11 24 1002en01 ...
Страница 80: ...Warning Displays 80 03 11 24 1008en01 ...
Страница 92: ...QC and Blood Controls 92 03 11 24 1156en01 ...
Страница 98: ...Maintenance Shut Down Transport 98 03 11 24 1011en01 ...
Страница 107: ......