WARNING:
Switch off the electrical supply to the pump and disconnect
the power cable from the electrical outlet before cleaning and inspection.
All equipment should be inspected. Any item that is visibly dirty should
be properly cleaned or removed. It is recommended that the system is
cleaned regularly and after each patient use.
SPECIFICATION
Control unit Specification
WARNING:
Medical electrical equipment needs special precautions
regarding EMC and needs to be installed according to the EMC
information provided. Careful consideration of this information is
essential when stacking or collocating equipment and when routing
cables and accessories.
WARNING:
RF mobile communications equipment can effect medical
electrical equipment.
Model Name
Comfort Glide Air Blower
Size in CM (L x W x H)
32.5 X 18 X 24.8
Weight
5 kgs
Air Hose
37mm dia x 1.5m length
Min/Max Pressure
20 ~ 30 mmHg ± 5mmHg
Flow Rate
≥ 2500 L/min
Rated Voltage
AC110 - 120V /60Hz
Max Current
10A
Fuse rating
F12AH 250V
Protection Type
Class II Type B
Environment (Temperature)
Operation: 15
o
C
to 35
o
C
(59
o
F
to 95
o
F
)
Storage: 5
o
C
to 60
o
C
(41
o
F
to 140
o
F
)
Environment (Humidity)
Operation: 30% to 75% non-condensing
Storage: 30% to 90% non-condensing
Filter Air
5 microns
Operation Atmospheric Pressure Range
700 hPa to 1060 hPa
Operation altitude
-1017 feet to 9,843 feet
Test Standard
IEC60601-1, IEC60601-1-2