background image

9

WARNING:

 Switch off the electrical supply to the pump and disconnect 

the power cable from the electrical outlet before cleaning and inspection.

All equipment should be inspected. Any item that is visibly dirty should 

be properly cleaned or removed. It is recommended that the system is 

cleaned regularly and after each patient use.

SPECIFICATION

Control unit Specification

WARNING:

 Medical electrical equipment needs special precautions 

regarding EMC and needs to be installed according to the EMC  

information provided. Careful consideration of this information is 

essential when stacking or collocating equipment and when routing 

cables and accessories.

WARNING:

 RF mobile communications equipment can effect medical 

electrical equipment.

Model Name

Comfort Glide Air Blower

Size in CM (L x W x H)

32.5 X 18 X 24.8

Weight

5 kgs

Air Hose

37mm dia x 1.5m length

Min/Max Pressure

20 ~ 30 mmHg ± 5mmHg

Flow Rate

≥ 2500 L/min

Rated Voltage

AC110 - 120V /60Hz

Max Current

10A

Fuse rating

F12AH 250V

Protection Type

Class II Type B

Environment (Temperature)

Operation: 15 

o

C

 to 35 

o

C

 (59 

o

F

 to 95 

o

F

)

Storage: 5 

o

C

 to 60 

o

C

 (41 

o

F

 to 140 

o

F

)

Environment (Humidity)

Operation: 30% to 75% non-condensing

Storage: 30% to 90% non-condensing

Filter Air

5 microns

Operation Atmospheric Pressure Range

700 hPa to 1060 hPa

Operation altitude

-1017 feet to 9,843 feet

Test Standard

IEC60601-1, IEC60601-1-2

Содержание ComfortGlide

Страница 1: ...USER MANUAL MSC60090 BLOWER Comfort Air GLIDE...

Страница 2: ...se 4 General safety 4 Device Feature 4 Inflating Comfort Glide Air Sheet 5 Maintenance Troubleshooting 6 Cleaning Disinfection Protocol 7 General Disassembly of Parts 7 Specification 9 Electromagnetic...

Страница 3: ...Blower Precaution Comfort Glide Air Blower To reduce the risk of burns electrocution fire or injury to persons 1 Use this product only for its intended use as described in this manual Do not use attac...

Страница 4: ...ans safely reposition and move patients The system has been designed to reduce risks and make patient repositioning and transfers easier and safer GENERAL SAFETY It is important to read the informatio...

Страница 5: ...will permit free movement of the hose CAUTION It is important to routinely inspect power cable to ensure it is not obstructed or causing a tripping hazard Check the power cable is not under strain or...

Страница 6: ...wer cord plug is securely connected and fits snugly into the outlet Secure plug connection and turn on Air Blower power switch Blown fuses in pump and plug Check for blown fuses in pump and plug Repla...

Страница 7: ...cility s guidelines for cleaning and disinfecting the Comfort Glide Air Blower Ensure that the power unit is off 2 Unplug the power cord from the wall outlet 3 Examine the surface of the Air Blower fo...

Страница 8: ...8 2 Changing filter a Unscrew 3 screws on the bottom cover b Open the filter cover and identify the filter c Replace the filter and assemble the parts in reverse order...

Страница 9: ...ng or collocating equipment and when routing cables and accessories WARNING RF mobile communications equipment can effect medical electrical equipment Model Name Comfort Glide Air Blower Size in CM L...

Страница 10: ...fort Glide Air Blower as rec ommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency...

Страница 11: ...s used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance for professional healthcare environment Electrostatic discharge ESD IEC 61000 4 2...

Страница 12: ...400 Pulse modulation b 18 Hz 1 8 0 3 27 27 450 430 470 GMRS 460 FRS 460 FM c 5 kHz deviation 1 kHz sine 2 0 3 28 28 710 704 787 LTE Band 13 17 Pulse modulation b 217 Hz 0 2 0 3 9 9 745 780 810 800 96...

Страница 13: ...calculated from the equation applicable to the frequency ofthe transmitter Recommendedseparationdistance d 1 2 d 1 2 80 MHz to 800 MHz d 2 3 800 MHz to 2 7 GHz Where P is the maximum output power rati...

Страница 14: ...transported another location Temperature limitations 5 o C to 60 o C 41 o F to 140 o F Relative Humidity 30 90 SYMBOLS USED Type B Applied part Refer to Instruction Manual Booklet Keep dry Waste Dispo...

Страница 15: ...at no charge This warranty excludes equipment damaged through shipping tampering improper maintenance carelessness accidents negligence misuse or which has been altered repaired or dismantled other th...

Страница 16: ...rademarks of Medline Industries Inc Made in Taiwan for Medline Industries Inc Three Lakes Drive Northfield IL 60093 USA RC19CMT Medline Industries Inc Three Lakes Drive Northfield IL 60093 USA Medline...

Отзывы: