Manufacturer’s declaration – electromagnetic emissions
The
Comfort Glide Air Blower
is intended for use in the electromagnetic environment (for profes-
sional healthcare) specified below.
The customer or the user of the
Comfort Glide Air Blower
should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment-guidance
(for professional healthcare environment)
RF emissions
CISPR 11
Group 1
The
Comfort Glide Air Blower
uses RF
energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A
The
Comfort Glide Air Blower
is suitable for use in
all establishments other than domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
N/A
Manufacturer’s declaration – electromagnetic immunity
The
Comfort Glide Air Blower
is intended for use in the electromagnetic environment
(for professional healthcare) specified below.
The customer or the user of the
Comfort Glide Air Blower
should assure that it is used in
such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment
-guidance (for professional
healthcare environment)
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,
±8 kV,±15 kV
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2kV for power
supply lines
Not applicable
Mains power quality should be
that of a typical professional
healthcare environment.
Surge
IEC 61000-4-5
± 0.5kV, ± 1kV
line(s) to line(s)
± 0.5kV, ± 1kV,
± 2kV line(s) to earth
± 0.5kV, ± 1kV
line(s) to line(s)
Not applicable
Mains power quality should be
that of a typical professional
healthcare environment.
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Voltage interruptions:
0 % UT;
250/300 cycle
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 30 cycles
Voltage interruptions:
0 % UT; 300 cycle
Mains power quality should be
that of a typical professional
healthcare environment. If the user
of the
Comfort Glide Air Blower
requires continued operation
during power mains interruptions,
it is recommended that the
Comfort Glide Air Blower
be
powered from an uninterruptible
power supply or a battery.