7
GENERAL DISASSEMBLY OF PARTS
Before disassembling any parts, ensure that the power unit is off and
unplug the power cord from the wall outlet.
1. Changing the air hose:
a. Use a flathead screwdriver to disconnect the prongs on the hose from
the blower unit.
b. Push the hose out to replace with a new one.
CLEANING & DISINFECTION PROTOCOL
1. Please follow your facility’s guidelines for cleaning and disinfecting
the
Comfort Glide Air Blower
. Ensure that the power unit is off.
2. Unplug the power cord from the wall outlet.
3. Examine the surface of the Air Blower for visible dirt and other
substances.
4. Refer to your Bio-Medical Engineering or Infection Prevention
Department for your facility’s protocol for cleaning and disinfecting.
5. Perform the following steps to clean the power unit and hose fittings:
- Wipe all controls, chassis and hose fittings with a damp cloth and a
mild
detergent.
- Air dry all treated surfaces.