10
User Manual
The installation will not be processed while the i700 wireless system is connected to a PC. Please be sure to disconnect the i700 wireless USB 3.0
cable from the PC before the installation.
2.2.3
Medit Scan for Clinics User Guide
Please refer to the User Guide of Medit Scan for Clinics: Medit Scan for Clinics > Menu > User Guide.
CAUTION
▪
Equipment maintenance should only be carried out by a Medit employee or Medit-Certified company or personnel.
▪
In general, users are not required to perform maintenance work on the i700 wireless system besides calibration, cleaning, and sterilization.
Preventive inspections and other regular maintenance are not required.
3. Maintenance
3.1
Calibration
Periodic calibration is required to produce precise 3D models. You should perform calibration when :
▪
The quality of the 3D model is not reliable or accurate when compared to previous results.
▪
Environmental conditions such as temperature have changed.
▪
The calibration period has expired.
You can set the calibration period in the Menu > Settings > Calibration Period (Days).
The calibration panel is a delicate component.
Do not touch the panel directly. Check the calibration panel if the calibration process is not performed properly. If the calibration panel is
contaminated, please contact your service provider.
The calibration panel is a delicate component.
You can set the calibration period via Menu > Settings > Calibration Period (Days). The default calibration period is 14 days.
3.1.1
How to Calibrate i700 wireless
①
Turn on the i700 wireless and launch the Medit Scan for Clinics.
②
Run the Calibration Wizard from Menu > Settings > Calibration.
③
Prepare the calibration tool and the i700 wireless handpiece.
④
Turn the dial of the calibration tool to position
1
.
⑤
Put the i700 wireless handpiece into the calibration tool.
⑥
Click “Next” to start the calibration process.
⑦
When the calibration tool is mounted properly in the correct position
1
, the system will automatically acquire the data.
⑧
When data acquisition is complete at position
1
, turn the dial to the next position.
⑨
Repeat the steps for positions
2
–
8
and the
LAST
position.
⑩
When data acquisition is complete at the
LAST
position, the system will automatically calculate and show the calibration results.
3.1.2
Auto Calibration Tool (Sold Separately)
The i700 wireless auto calibration tool accessory can be purchased separately. This handy auto calibration tool will automatically perform calibration by
calibrating the i700 wireless handpiece without having to turn the calibration dial. Please refer to Medit Scan for Clinics for details.
3.2
Cleaning, Disinfection, Sterilization Procedure
3.2.1
Reusable Tip – Sterilization
The reusable tip is the part that is inserted into the patient’s mouth during scanning. The tip is reusable for a limited number of times. The tip needs to
be cleaned and sterilized between patients to avoid cross-contamination.
▪
The tip should be cleaned manually using a disinfecting solution. After cleaning and disinfecting, inspect the mirror inside the tip to
ensure there aren’t any stains or smudges.
▪
Repeat the cleaning and disinfection process if necessary. Carefully dry the mirror using a paper towel.
▪
Insert the tip into a paper sterilization pouch and seal it, making sure that it is airtight. Use either a self-adhesive or heat-sealed pouch.
▪
Sterilize the wrapped tip in an autoclave with the following conditions:
▪
Sterilize for 30 minutes at 121°C (249.8°F) at gravity type and dry for 15 minutes.
▪
Sterilize for 10 minutes at 135°C (275°F) at gravity type and dry for 30 minutes.
▪
Sterilize for 4 minutes at 134°C (273.2°F) at pre-vacuum type and dry for 20 minutes.
▪
Use an autoclave program that dries the wrapped tip before opening the autoclave.
▪
Scanner tips can be re-sterilized up to 150 times and thereafter must be disposed of as described in the disposal section.
▪
Autoclave times and temperatures may vary depending on the autoclave type and manufacturer. For this reason, it may not be able to meet the
maximum number of times. Please refer to the user’s manual of the autoclave manufacturer you are using to determine whether the required
conditions are met.
3.2.2
Reusable Tip – Cleaning and Disinfection
▪
Clean the tip immediately after use with soap water and a brush. We recommend using a mild dishwashing liquid. Make sure the mirror of the
tip is completely clean and stain-free after cleaning. If the mirror appears stained or foggy, repeat the cleaning process and rinse thoroughly with
water. Dry the mirror carefully with a paper towel.
▪
Clean with Caviwipes according to the following conditions. Please refer to the Caviwipes instruction manual for proper use.
▪
Caviwipes : disinfect for 3 minutes and dry for 5 minutes.
▪
Caviwipes-1 : disinfect for 1 minute and dry for 5 minutes.
▪
Caviwipes-2 : disinfect for 2 minutes and dry for 5 minutes.
▪
Disinfect the tip using Wavicide-01 for 45 to 60 minutes. Rinse the tip thoroughly. Please refer to the Wavicide-01 Solution instruction
manual for proper use.
▪
Remove the tip from the used solution and rinse thoroughly after cleaning and sterilization.
▪
Use a sterilized and non-abrasive cloth to dry the mirror and the tip gently.
Содержание i700
Страница 1: ...Revision 1 January 2022...
Страница 2: ......
Страница 3: ...CONTENTS English Fran ais Deutsch Italiano Portugu s Espa ol 02 20 38 54 72 90 108 126 144 162...
Страница 4: ...English...
Страница 22: ......
Страница 40: ......
Страница 56: ...Fran ais...
Страница 74: ...Deutsch...
Страница 92: ...Italiano...
Страница 110: ...Portugu s...
Страница 128: ......
Страница 134: ...132 User Manual 2 30 1 7 5 i700 wireless i700 wireless i700 wireless 1 7 6 1 7 7 i700 wireless i700 wireless...
Страница 146: ...Espa ol...
Страница 164: ......
Страница 178: ...176 NOTE...
Страница 179: ...177 NOTE...
Страница 180: ...178 NOTE...
Страница 181: ...179 NOTE...
Страница 182: ...180 NOTE...
Страница 183: ...181 NOTE...
Страница 184: ...182 NOTE...
Страница 185: ...183 NOTE...
Страница 186: ...184 NOTE...
Страница 187: ...185 NOTE...
Страница 188: ...186 NOTE...
Страница 189: ...187 NOTE...
Страница 190: ...188 NOTE...
Страница 191: ...189 NOTE...
Страница 192: ...190 NOTE...
Страница 193: ...191 NOTE...
Страница 194: ...192 NOTE...
Страница 195: ...193 NOTE...
Страница 196: ...194 NOTE...
Страница 197: ...195 NOTE...
Страница 198: ...196 NOTE...
Страница 199: ...197 NOTE...
Страница 200: ...198 NOTE...
Страница 201: ...199 NOTE...
Страница 202: ...200 NOTE...
Страница 203: ...201 NOTE...
Страница 204: ...202 NOTE...
Страница 205: ...203 NOTE...
Страница 206: ...204 NOTE...
Страница 207: ...205 NOTE...
Страница 208: ...206 NOTE...
Страница 209: ...207 NOTE...
Страница 210: ...208 NOTE...
Страница 211: ...209 NOTE...
Страница 212: ...210 NOTE...
Страница 213: ...211 NOTE...
Страница 214: ...212 NOTE...
Страница 215: ...213 NOTE...
Страница 216: ...214 NOTE...
Страница 217: ...215 NOTE...
Страница 218: ...216 NOTE...
Страница 219: ...217 NOTE...
Страница 220: ...218 NOTE...
Страница 221: ...219 NOTE...
Страница 222: ...220 NOTE...
Страница 223: ...221 NOTE...
Страница 224: ...222 NOTE...