II.
Functional check of the handrails
The handrails must be checked in accordance with DIN EN 60601-2-52.
Check that the locking mechanism of the handrails is working correctly and that there
are no visible deformations or signs of wear on the handrails.
Check that the prescribed distances are maintained even with the rails down.
All the measurements of the distances of the side rails should be made with the bed in
a horizontal position, since the given dimensions while the bed is placed in another way
do not show less favorable conditions.
The measurement of A and D must be carried out using a cone defined in DIN EN
60601-2-52. The force used for the measurement of A must be at least 250 N
Figure N. 21- Functional check of side rails
Table N. 14-
Dimensions
Letra
Descripción
Dimensión
A
x
The smallest dimension between the elements inside the perimeter of
the railing in its raised / locked positions or the perimeters created
between the railing and the fixed parts of the bed
< 120 mm
B
Distance between mattress rest area and lowest point of the side rail,
out of the support of the side rail
< 60 mm
C
1
Distance between headboard and side rail
< 60 mm
C
2,3
Distance between Split side rails in their upper position and distance
between side rail and footboard
< 60 mm o > 318 mm.
D
Area between side rail and mattress
El cono de 120 mm debe
hundir como máximo 60
mm la superficie del
colchón sin aplicar presión