7
Vacuum levels in Vcare
®
system should be tailored to each specific wound and
adjusted according to patient's clinical condition. General guidelines and
recommended vacuum levels are shown in the table below.
Inexperienced personnel is advised to always use the cyclic-continuous mode or
consult a trained physician regarding the treatment settings before using the Vcare
®
system.
When applying external vacuum - make sure that the vacuum level supplied from
the external vacuum source is 50 mmHg above the desired working pressure but
should not exceed 200 mmHg.
It is recommended to connect the unit to an external vacuum source by a pressure
regulator.
The external vacuum tube must be disconnected from the unit when external
vacuum is not in use.
When working with the unit’s internal pump, make sure that the external vacuum
tube is not connected to the unit.
The device must be operated in a quite environment with background noise of no
more than 45-50dB
A
portable modality of utilization enables via a mobile stand.
IV (intravenous) medication administration by IV pump can be integrated with
Vcare
®
treatment. The pump shall be hanged on IV pole that structured in the
mobile stand or being placed on a flat surface that is higher than the Vcare
®
unit.
Warnings regarding clinical application of the Vcare
®
system
Bleeding and infection should be anticipated during the use of RNPT.
When bleeding risk is anticipated to be high (early following trauma or surgery),
working negative pressure levels should be set as low as possible and should not
exceed a maximum of 80 mmHg.
While working on stiff surfaces, as in clinical cases of low- perfused tissues like
PVD, diabetic, traumatic wounds, low vacuum is essential in order to prevent
tissue ischemia.
Maximal Flow Rate Setting: Maximal flow rate from the wound through the tubing
to the collection canister is pre-set as default to be up to 100 ml/hr. In case that the
flow exceeds this limit, an alarm is activated and the suction apparatus shuts