26
If the chosen value is YES, the previous set values of the following parameters
will be displayed, one after the other, on the second text line:
•
Working neg. Pressure.
•
Upper / Lower Limit.
•
Max. Flow.
•
Work / Pause Time (displayed only if the previous Mode of Operation is
Intermittent).
After viewing the previous settings, the following options are available:
•
Press ↓ to go to Stand-by screen.
•
Press FUNC to adjust treatment settings.
Pressing FUNC will transfer the user to the next function in the menu. In this
case, the flickering values will be the values chosen in the previous treatment,
and the user will be able to change them.
5.1.3. Mode of Operation
When entering the Mode of Operation function, the following screen is
displayed:
Insert printout
The default mode of operation, Cyclic-Continuous mode, is flickering.
There are two optional modes of operation: Intermittent or Cyclic- Continuous.
Use the up and down scroll arrow buttons in the control panel to change the
flickering mode to the required mode of operation. The guiding messages in
the second line change according to the selected mode.
Press OP/SET to finalize your selection. The value stops flickering.
Press ↓ to continue to the next parameter.
See Clinical guidelines and Vacuum Operation Modes sections before
choosing the Mode of Operation
.
5.1.3.1.
Cyclic-Continuous Working Negative Pressure
The following screen is displayed:
Insert printout
The default value of cyclic-continuous working neg. pressure is flickering.