46
the selection by pressing the OP/SET button.
The user needs to follow the
instruction in user manual section 3.1-3.3. Only after conducting all the
required
procedures the user need to press the "Done" botton. The default selection of this
alert is “Postpone”.
3.1.
Device Service and Maintenance
Description: in order to maintain the device performance and effective operation, the
Vcare
®
unit should go through routine maintenance inspection. Routine
maintenance is required after 5000 hours or following one year of engine operation
(whichever comes first).
Alarm signals: none.
Additional signals: general warning status (symbol no. 7) from screen symbols
description table (the signals will appear only after 4500 hours of operation).
Information sentences: First line:
“Postpone Done”, Second line: (1) “Maintenance
required”, (2) “Refer to user manual”:
To resolve this alert:
Contact IVT Medical Ltd. for a certified technical personnel inspection of:
1.
Internal and external fuses inspection
2.
Battery voltage measurement of maximal voltage and inspection for possible leakage
and corrosion
3.
Replacing the internal pump
4.
Inner hydro-gel filter replacement
5.
Fluid trapped replacement
6.
Clear PVC tubes inspection and replacement if indicated
7.
Fan cleaning
8.
Tightening of loose screws
9.
Alarm Buzzer and LEDs inspection
10.
Pressure sensors validation
11.
Screws on the outer shell should be inspected for fatigue and replaced if necessary.
For further guidance and maintenance instructions, refer to Maintenance chapter in
this manual.