44
Additional signals: general warning status (symbol no. 7) and canister fill status-
indicates the canister fill level (symbol no. 10) from screen symbols description table.
Information sentences: First line:
“Full canister”, Second line” (1) “Turn system off (2)
replace canister and” (3) “Re-activate treatment” (where (1), (2) and (3) are presented
one after the other for 5 seconds each).
Additional signal: symbol number 9 appears on the screen.
To resolve this alarm:
•
Stop the vacuum operation.
•
Turn the Vcare
®
unit off.
•
Replace the collection canister as instructed in replacing the collection canister
section in this manual.
•
Restart the device.
In case this alarm is repeated after following these instructions, contact your service
provider.
Vacuum operation stops automatically when the fluids level in the canister is 700 ml.
In case this alarm is repeated, contact your service provider.
For further information regarding the alarm system, please refer to the alarm appendix.
In case of mechanical or electronic failure of the device and for any query or
improvement suggestion contact your service provider.
2.
Alert conditions
2.1.
High Vacuum alert
Description: the applied neg. pressure is above the upper limit.
Alarm signals: none.
Additional signals: general warning status (symbol no. 7) from screen symbols
description table.
Information sentences: Second line:
(1)”Caution: Pressure is” (2) “above upper limit”
(where (1) and (2) are presented one after the other for 5 seconds each).
To resolve this alert:
•
Stop the vacuum operation.
•
Turn the Vcare
®
unit off.
•
If connected to an external vacuum source, reduce the pressure from it