45
•
Restart the device.
In case this alarm is repeated after following these instructions, contact your service
provider.
In order to confirm that the preset vacuum is actually applied to the wound, it is
recommended that together with the treatment operational setting, the care provider
should inspect the sponge to be evenly condensed/ squeezed.
Warning: failure to respond to this alert may lead to High Vacuum alarm
condition.
2.2.
Low Vacuum alert
Description: the applied neg. pressure is below the lower limit.
Note: as vacuum operation is initiated, the low vacuum alert will be displayed on the
screen until the working negative pressure stabilizes on the desired level, unless it
cannot be stabilized due to leakage.
Alarm signals: none.
Additional signals: general warning status (symbol no. 7) from screen symbols
description table.
Information sentences: Second line:
(1)”Caution: Pressure is” (2) “below lower limit”.
To resolve this alert:
•
Inspect the sponge, drape stripes, tubing and collection canister for possible
leakage.
•
Make sure that the canister is positioned properly in the canister socket.
•
Make sure that the pressure from the external vacuum source is above the
working pressure.
In case this alarm is repeated after following these instructions, contact your service
provider.
Warning: failure to respond to this alert may lead to Low pressure alarm
condition.
3.
Maintenance and Error alerts
The maintenance and error alerts are displayed, when necessary, following the self
test screen. The user can choose to postpone the alert to a later stage (Postpone)
or to resolve it immediately (Done) by using the scroll arrow buttons and finalizing