The available working neg. pressure and default value are
determined according to the Risk of Bleeding (see the table below
for detailed default working neg. pressure and displayed values).
In addition, the working neg. pressure can be linearly adjusted
within the range of 30-200 mmHg, and is limited in accordance to
the Risk of Bleeding evaluation, as described in the table below.
Risk for Bleeding
Default Working
Pressure
[mmHg]
Displayed Pressure
Values
[mmHg]
Working Pressure
Range
[mmHg]
High
60
40, 50, 60
30-65
Moderate
70
60, 70, 80
30-90
Low
80
60, 80, 100
No limit (30-200)
Use the up and down scroll arrow buttons in the control panel to
change the flickering value and determine the required Working
neg. Pressure.
Press OP/SET to finalize your selection. The value stops
flickering.
For advanced pressure setting, press FUNC shortly. The working
neg. pressure value will be displayed in [a].
Use the up and down scroll arrow buttons in the control panel to
change the displayed value and determine the required Working
neg. Pressure. Press OP/SET to finalize your selection. The
displayed value stops flickering.
Pre
ss ↓ to continue to the next parameter.
5.1.3.2.2. Lower Limit
When entering the Lower Limit function, the following screen is
displayed:
Insert printout
The default lower limit value is flickering.