MediCap
®
USB200 User Guide – English
7 of 20
USB200H-UG01-EN-140120-MC
Typical Connections
Important Note:
There is a
very brief
delay (0.06 sec.) between the time when the video signal enters
the USB200 and when it is displayed on a monitor connected to the USB200. If this delay is not suitable
for your medical environment, do not connect your monitor to the USB200. Instead, connect your
monitor directly to your video source; then connect the USB200 to the secondary output on your video
source or use a “Y-Splitter” (not included).
The unit is for exclusive interconnection with IEC60601 certified equipment in the patient environment and
IEC 60XXX certified equipment outside of the patient environment.
A certified power supply cord set must be used with this equipment. The relevant national installation and/or
equipment regulations shall be considered. The power supply cord shall be not less robust than an ordinary tough
rubber-sheathed flexible cord according to IEC 60245 (designation 53) or ordinary polyvinyl chloride sheathed
flexible cord according to IEC 60227 (designation 53).
The power cord set shall have a power supply cord with conductors providing a cross-sectional area of 3 x
0.75mm
2
minimum.