MedaCure Comfort Zone CZ36 Скачать руководство пользователя страница 3

         

   

2

 

Class I Medical Device 

according to Directive 93/42/EEC 

and further modifications 

MODEL # & DESCRIPTION

 

CZ36 & CZ36-BIP (WITH BUILT-IN PERIMITER) PUMP & MATTRESS - 36” WIDTH, 8” CELL-ON-CELL HEIGHT

INTRODUCTION

 

Thank you for purchasing the Comfort Zone Mattress with Alternating Pressure and Low Air Loss

 

by 

MedaCure.

 This user’s manual provides suggestions on how to correctly use the product and gives   

valuable advice regarding your safety. Please read through the manual carefully before using the product.

 

Should you have any questions, please contact your dealer for advice and assistance. 

 

INTENDED USE

 

The Comfort Zone mattress is designed for bed sore and wound care therapy treatment and prevention, which may occur 
during an extended hospital stay and nursing home /long term care environment.

 

CAUTION! 

Do not use the product for any purpose not indicated in this manual.

Only qualified personnel trained in the treatment and prevention of bedsores should operate this device.

MedaCure Inc. declines all liability for any consequences resulting from incorrect use of this product and

from unauthorized alteration to the frame of the product.

The manufacturer reserves the right to change the information contained in this document without prior

notice

For patients

When the patient is in a sitting position, check to see that the patient is 

not bottoming out – the pressure setting is inadequate (or too low) – and 

the patient is hitting the bed frame. 

See the figure at right. If there is an incident of bottoming out, simply 

adjust the pressure range a little higher.

HAND CHECK

Slide one hand between the patient’s sacral region and the air mattress 

to feel whether the pressure is properly adjusted. The optimal clearance 

range is 1” to 1½” (25 to 40 mm). 

NOTE

The hand check procedure is recommended by the Agency for Health 

Care Policy and Research.

Содержание Comfort Zone CZ36

Страница 1: ...User Manual FOR MODEL CZ36 CZ36 BIP 8 REPLACEMENT CELL ON CELL MATTRESS 2016 MADE IN CHINA...

Страница 2: ...ORE USE page 6 USE WARNINGS page 6 7 HOW TO USE page 7 HOW TO USE THE CPR DEFLATOR INFLATOR page 7 8 STORAGE page 9 MAINTENANCE page 9 CLEANING AND DISINFECTION page 9 CONDITIONS OF DISPOSAL page 9 Ge...

Страница 3: ...ct for any purpose not indicated in this manual Only qualified personnel trained in the treatment and prevention of bedsores should operate this device MedaCure Inc declines all liability for any cons...

Страница 4: ...nt otherwise injury or damage may occur 2 Consult with physician or therapist to determine the correct adjustment and the correct use of this device 3 Keep the packed kit away from heat sources 4 SERV...

Страница 5: ...ernate keeps mattress firm and always inflated d SEAT FUNCTION To increase pressure in each cell when mattress is in the sitting position COMFORT LOW PRESSURE To increase or decrease airflow for a sof...

Страница 6: ...r 5 Fuse Back 8 Filter 9 Main filter 10 Protective cap of filter 11 Hook 1 Pillow 2 Cover 3 Fixing belt 4 Low air loss holes 5 Male quick connector 6 Protective cap of connector 7 Air tubing 8 Patente...

Страница 7: ...e of a sudden power outage The mattress will temporarily retain air even when the pump is turned off or during a power outage Please note this only temporary The cover sheet protects the cells against...

Страница 8: ...plug the control unit into the ground power supply 7 Do not use the product in case of the following 1 the plug or the cable is damaged 2 the system sounds abnormal or vibrates intensely 3 the control...

Страница 9: ...ng 4 Align the air holes as shown to deflate 1 This mattress includes a Quick Inflate Pump and nozzle attachment 2 Attach the nozzle to the pump as shown 3 Align the air holes on the CPR Insert the Qu...

Страница 10: ...t by bringing it to a specific recycling point for electric and electronic equipment or others that provide this service SPARE PARTS ACCESSORIES For spare parts and accessories please contact MedaCure...

Страница 11: ...emissions CISPR 11 Group 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment...

Страница 12: ...to line s 2kV line s to earth 1kV differential mode 2kV differential d Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage...

Страница 13: ...ngths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment...

Страница 14: ...ower of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73...

Страница 15: ...peration 10 40 Storage 15 50 Shipping 15 70 Environment humidity Operation 10 90 non condensing Storage 10 90 non condensing Shipping 10 90 non condensing Atmospheric Pressure Operation 700 1013 25hPa...

Страница 16: ...alling Patient must be under the constant watch of a caregiver to monitor the patient s movements When the patient needs to transfer out of bed each side can be deflated individually to allow the pati...