Class I Medical Device
according to Directive 93/42/EEC
and further modifications
MODEL # & DESCRIPTION
CZ36 & CZ36-BIP (WITH BUILT-IN PERIMITER) PUMP & MATTRESS - 36” WIDTH, 8” CELL-ON-CELL HEIGHT
INTRODUCTION
Thank you for purchasing the Comfort Zone Mattress with Alternating Pressure and Low Air Loss
by
MedaCure.
This user’s manual provides suggestions on how to correctly use the product and gives
valuable advice regarding your safety. Please read through the manual carefully before using the product.
Should you have any questions, please contact your dealer for advice and assistance.
INTENDED USE
The Comfort Zone mattress is designed for bed sore and wound care therapy treatment and prevention, which may occur
during an extended hospital stay and nursing home /long term care environment.
CAUTION!
•
Do not use the product for any purpose not indicated in this manual.
•
Only qualified personnel trained in the treatment and prevention of bedsores should operate this device.
•
MedaCure Inc. declines all liability for any consequences resulting from incorrect use of this product and
from unauthorized alteration to the frame of the product.
•
The manufacturer reserves the right to change the information contained in this document without prior
notice
For patients
When the patient is in a sitting position, check to see that the patient is
not bottoming out – the pressure setting is inadequate (or too low) – and
the patient is hitting the bed frame.
See the figure at right. If there is an incident of bottoming out, simply
adjust the pressure range a little higher.
HAND CHECK
Slide one hand between the patient’s sacral region and the air mattress
to feel whether the pressure is properly adjusted. The optimal clearance
range is 1” to 1½” (25 to 40 mm).
NOTE
The hand check procedure is recommended by the Agency for Health
Care Policy and Research.