ELECTROMAGNETIC COMPATIBILITY DECLARATION 

Guidance and manufacturer’s declaration-electromagnetic immunity 

The device is intended for use in the electromagnetic environment specified below. 

The customer or the user of the device should assure that it is used in such an environment. 

Immunity test 

IEC 60601 

test level 

 Compliance level 

Electromagnetic 

environment-guidance 

Electrostatic discharge 
(ESD) 

IEC 61000-4-2 

+ 6 kV contact

+ 8 kV air

+ 6 kV contact

+ 8 kV air

Floors should be wood, concrete 
or ceramic tile. If floors are covered 
with synthetic material, the 
relative humidity should be at least 
30% 

Electrical  
fast transient/burst 

IEC 61000-4-4 

+ 2kV for power
supply lines

+ 1kV for input/output

lines

+ 2kV for power
supply lines

Not applicable 

Mains power quality should be that of 
a typical commercial or 
hospital environment. 

Surge IEC 61000-4-5 

+ 1kV line(s) to line(s)

+ 2kV line(s) to earth

+ 1kV differential
mode

+ 2kV differential

d

Mains power quality should be that 
of a typical commercial or 
hospital environment. 

Voltage Dips, short 

interruptions and voltage 

variations on power 

supply input lines IEC 

61000-4-11 

<5% UT(>95% dip in

UT) for 0,5 cycle 

40% UT(60% dip in 

UT) for 5 cycles 

70% UT(30% dip in 

UT) for 25 cycles 

<5% UT(>95% dip in

UT) for 5 s 

<5% UT(>95% dip in

UT) for 0,5 cycle 

40% UT(60% dip in 

UT) for 5 cycles 

70% UT(30% dip in 

UT) for 25 cycles 

<5% UT(>95% dip in

UT) for 5 s 

Mains power quality should be that of 
a typical commercial or 
hospital environment. If the user 
of the device requires continued 
operation during power 
mains interruptions, it is recommended
that the device be powered from 
an uninterruptible power supply 
or a battery. 

Power frequency(50/60 
Hz) magnetic field IEC 
61000-4-8 

3 A/m 

3 A/m 

The device power 

frequency magnetic fields should 

be at levels characteristic of a 

typical location in a typical 

commercial or hospital 

environment. 

NOTE 

UT is the a.c. mains voltage prior to application of the test level. 

Содержание Comfort Zone CZ36

Страница 1: ...User Manual FOR MODEL CZ36 CZ36 BIP 8 REPLACEMENT CELL ON CELL MATTRESS 2016 MADE IN CHINA...

Страница 2: ...ORE USE page 6 USE WARNINGS page 6 7 HOW TO USE page 7 HOW TO USE THE CPR DEFLATOR INFLATOR page 7 8 STORAGE page 9 MAINTENANCE page 9 CLEANING AND DISINFECTION page 9 CONDITIONS OF DISPOSAL page 9 Ge...

Страница 3: ...ct for any purpose not indicated in this manual Only qualified personnel trained in the treatment and prevention of bedsores should operate this device MedaCure Inc declines all liability for any cons...

Страница 4: ...nt otherwise injury or damage may occur 2 Consult with physician or therapist to determine the correct adjustment and the correct use of this device 3 Keep the packed kit away from heat sources 4 SERV...

Страница 5: ...ernate keeps mattress firm and always inflated d SEAT FUNCTION To increase pressure in each cell when mattress is in the sitting position COMFORT LOW PRESSURE To increase or decrease airflow for a sof...

Страница 6: ...r 5 Fuse Back 8 Filter 9 Main filter 10 Protective cap of filter 11 Hook 1 Pillow 2 Cover 3 Fixing belt 4 Low air loss holes 5 Male quick connector 6 Protective cap of connector 7 Air tubing 8 Patente...

Страница 7: ...e of a sudden power outage The mattress will temporarily retain air even when the pump is turned off or during a power outage Please note this only temporary The cover sheet protects the cells against...

Страница 8: ...plug the control unit into the ground power supply 7 Do not use the product in case of the following 1 the plug or the cable is damaged 2 the system sounds abnormal or vibrates intensely 3 the control...

Страница 9: ...ng 4 Align the air holes as shown to deflate 1 This mattress includes a Quick Inflate Pump and nozzle attachment 2 Attach the nozzle to the pump as shown 3 Align the air holes on the CPR Insert the Qu...

Страница 10: ...t by bringing it to a specific recycling point for electric and electronic equipment or others that provide this service SPARE PARTS ACCESSORIES For spare parts and accessories please contact MedaCure...

Страница 11: ...emissions CISPR 11 Group 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment...

Страница 12: ...to line s 2kV line s to earth 1kV differential mode 2kV differential d Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage...

Страница 13: ...ngths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment...

Страница 14: ...ower of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73...

Страница 15: ...peration 10 40 Storage 15 50 Shipping 15 70 Environment humidity Operation 10 90 non condensing Storage 10 90 non condensing Shipping 10 90 non condensing Atmospheric Pressure Operation 700 1013 25hPa...

Страница 16: ...alling Patient must be under the constant watch of a caregiver to monitor the patient s movements When the patient needs to transfer out of bed each side can be deflated individually to allow the pati...