ELECTROMAGNETIC COMPATIBILITY DECLARATION 

  Guidance and manufacturer’s declaration-electromagnetic immunity 

The CZ36 & CZ36-BIP is intended for use in the electromagnetic environment specified below. 

The customer or the user of the CZ36 & CA36-BIP should ensure that it is used in such an environment.

  Immunity test 

IEC 60601 test 

level 

Compliance 

level 

Electromagnetic 

environment-guidance 

Conducted RF 
IEC 61000-4-6 
Radiated RF 
IEC 61000-4-3 

3 Vrms 
150 KHz to 80 
MHz 
3 V/m 
80MHz to 2,5 GHz

3 Vrms 
3 V/m 

Portable and mobile RF communications 
equipment 
should be used no closer to any part of 
the CZ36 & CZ36-BIP 
including cables, than the recommended 
separation distance calculated from the 
equation applicable to the frequency of 
the transmitter. 

Recommended separation distance: 

d = 1,2 

d = 1,2 

P

 80MHz to 800 MHz 

d = 2,3 

P

 800MHz to 2,5 GHz 

Where

 P

 is the maximum output power 

rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and

 d

 is the recommended 

separation distance in meters (m). 

(a) Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
(b) Interference may occur in the vicinity
of equipment marked with the following
symbol:

NOTE1      At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE2      These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 
structures, objects and people. 
a    Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 

amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 
field strength in the location in which the CZ36 & CZ36-BIP

 

is used exceeds the applicable RF compliance level above, the CZ36 & 

CZ36-BIP

 

should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, 

such as re-orienting or relocating the CZ36 & CZ36-BIP 

b    Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

Содержание Comfort Zone CZ36

Страница 1: ...User Manual FOR MODEL CZ36 CZ36 BIP 8 REPLACEMENT CELL ON CELL MATTRESS 2016 MADE IN CHINA...

Страница 2: ...ORE USE page 6 USE WARNINGS page 6 7 HOW TO USE page 7 HOW TO USE THE CPR DEFLATOR INFLATOR page 7 8 STORAGE page 9 MAINTENANCE page 9 CLEANING AND DISINFECTION page 9 CONDITIONS OF DISPOSAL page 9 Ge...

Страница 3: ...ct for any purpose not indicated in this manual Only qualified personnel trained in the treatment and prevention of bedsores should operate this device MedaCure Inc declines all liability for any cons...

Страница 4: ...nt otherwise injury or damage may occur 2 Consult with physician or therapist to determine the correct adjustment and the correct use of this device 3 Keep the packed kit away from heat sources 4 SERV...

Страница 5: ...ernate keeps mattress firm and always inflated d SEAT FUNCTION To increase pressure in each cell when mattress is in the sitting position COMFORT LOW PRESSURE To increase or decrease airflow for a sof...

Страница 6: ...r 5 Fuse Back 8 Filter 9 Main filter 10 Protective cap of filter 11 Hook 1 Pillow 2 Cover 3 Fixing belt 4 Low air loss holes 5 Male quick connector 6 Protective cap of connector 7 Air tubing 8 Patente...

Страница 7: ...e of a sudden power outage The mattress will temporarily retain air even when the pump is turned off or during a power outage Please note this only temporary The cover sheet protects the cells against...

Страница 8: ...plug the control unit into the ground power supply 7 Do not use the product in case of the following 1 the plug or the cable is damaged 2 the system sounds abnormal or vibrates intensely 3 the control...

Страница 9: ...ng 4 Align the air holes as shown to deflate 1 This mattress includes a Quick Inflate Pump and nozzle attachment 2 Attach the nozzle to the pump as shown 3 Align the air holes on the CPR Insert the Qu...

Страница 10: ...t by bringing it to a specific recycling point for electric and electronic equipment or others that provide this service SPARE PARTS ACCESSORIES For spare parts and accessories please contact MedaCure...

Страница 11: ...emissions CISPR 11 Group 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment...

Страница 12: ...to line s 2kV line s to earth 1kV differential mode 2kV differential d Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage...

Страница 13: ...ngths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment...

Страница 14: ...ower of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73...

Страница 15: ...peration 10 40 Storage 15 50 Shipping 15 70 Environment humidity Operation 10 90 non condensing Storage 10 90 non condensing Shipping 10 90 non condensing Atmospheric Pressure Operation 700 1013 25hPa...

Страница 16: ...alling Patient must be under the constant watch of a caregiver to monitor the patient s movements When the patient needs to transfer out of bed each side can be deflated individually to allow the pati...

Отзывы: