Initial Setup
License Key
A VectorPro Lite license key is required to run
a Vortex-
dV
test stand (fitted with an AFTI) and VectorPro software.
Locating the Stand
The test stand should be positioned on a suitable, level, stable work surface.
Mains Power Supply
Vortex
-dV
test stands can be used on 110–120 or 220–240 Vac 50-60 Hz supplies. The rear fuse carrier is set for
your local requirement but is reversible. Should you replace a fuse, ensure the correct local voltage is selected.
The voltage that is selected is indicated by which the arrow is pointing to the white line located at the bottom of
the device. This is illustrated in the image below, shown within the red circle:
Fuse Specification
A Vortex
-dV
test stand uses two 2A - Time-delay (T), 5 x 20mm fuses. If replacing a blown fuse only replace the
fuse on the active side of the inlet filter with the fuse specified above, or equivalent.
If you are in doubt, please contact your local Mecmesin support agent for more information.
Attaching the Torque Sensor
The torque cell attaches to the crosshead using the four supplied M6 caphead screws. To fit the torque cell, first,
ensure that the cable exit is towards the side that you plan to fit the AFTI.
Then, align the transducer with the mounting plate and secure using one the M6 bolts, if needed have one person
hold the transducer while the other fits the screws. Make sure the transducer is seated centrally before fulling
tightening all of the screws.
10
Содержание Vortex-dV
Страница 2: ...1 Vortex dV Motorised Torque Test Stand Operating Manual 431 468 April 2020 2 ...
Страница 15: ...Now press RESET ENTER on PORT Step 5 Set TX UNITS to ON press RESET ENTER to continue Step 6 15 ...
Страница 18: ...Step 10 Ensure the TX threshold is set to 0 press REST ENTER to continue Step 11 18 ...
Страница 35: ...10 Dimensions Note Vortex dV Dimensions mm inch 35 ...