Half Cycle
A half-cycle test is to a displacement relative to tared zero. A cycle starts when the
crosshead is at the first displacement position and ends back at the second position.
Setting
Options
Cycle Count
0-9999
Up Speed
0-1200mm/min
Down Speed
0-1200mm/min
Upper Displacement
A positive displacement is above tared zero and a negative value is
below. Available displacement depends on the test setup.
Lower Displacement
A positive displacement is above tared zero and a negative value is
below.
Start Direction
Choose whether the test direction is upward or downward.
Move to Start
Select if the test moves to the start position first.
Example
Upper displacement:
+30 mm
Lower displacement:
–
20 mm
Initial stroke
:
Down
Move to Start:
Yes
Unless already at 30 mm above tared zero, the crosshead will travel to that point and
then move to 20 mm below tared zero, and stop.