MultiTest
-dV
Operating Manual
Mecmesin
35
Setting
Options
Start Direction
Choose whether the test direction is upward or downward.
Test Sub-Type
Select the test sub-type (Options are listed above).
Limit Force
Limit and Cyclic Tests Only -
Enter the target load for the test.
Return Force
Cyclic Tests Only -
Enter the start load for the test.
Cycle Count
Cyclic Tests Only -
Enter the number of cycles to be completed.
Break Threshold
Move to Break Test Only –
Enter the percentage drop from current
maximum load recorded, to activate the break detection.
Example:
Current load maximum reading is 1000N, with 80% setting
the load drop must reach 200N before break detection activates.
Min Break Threshold
Move to Break Test Only -
Enter the minimum break threshold
(value of load that the test value must rise above for a break
condition to be detected).
AFG must be connected and ON to set this parameter
Examples
Force Limit
Up Speed:
100 mm/min
Down Speed:
100 mm/min
Start Direction:
Down
Test Sub-Type:
Force Limit
Limit Force:
50 N
The crosshead moves down at 100 mm/min until the applied load is 50 N from tared
zero, once the limit force is reached the test stops.
Содержание MultiTest 0.5-dV
Страница 1: ...MultiTest dV Tension and Compression Test Stand Operating Manual 431 459 04 February 2019 ...
Страница 44: ...44 Mecmesin MultiTest dV Operating Manual MultiTest dV 1 0kN Dimensions Note The above drawing is not to scale ...
Страница 45: ...MultiTest dV Operating Manual Mecmesin 45 MultiTest dV 0 5kN Dimensions Note The above drawing is not to scale ...
Страница 46: ...46 Mecmesin MultiTest dV Operating Manual 11 Declaration of Conformity ...
Страница 51: ...MultiTest dV Operating Manual Mecmesin 51 14 Operator Notes ...