40
Mecmesin
MultiTest
-dV
Operating Manual
7.2.
MultiTest
-dV
Settings -
dV
(u) Operation
The settings listed below are for a MultiTest
-dV
being used in
dV
(u) mode, for standard
dV
operation see
7.1.
All settings are made by moving the selection marker to the required item or digit and
confirming with the tick button or using the central scroll wheel button.
Settings: Jog settings
Within the jog settings menu, you can configure the jog speed limits while in jog mode.
Below is a detailed breakdown of each setting and the options available for each setting.
Setting
Action
Range
Up Speed
Configure the jog speed in an upward motion
0.1 to 1200
mm/min
0.004 to 47.24
Inch/min
Down Speed
Configure the jog speed in a downward motion
0.1 to 1200
mm/min
0.004 to 47.24
Inch/min
Tension Limit
Configure the tensile force limit while in jog mode
0 to +200 N
Compression Limit
Configure the compressive force limit while in jog mode
0 to -200 N
Settings: PIN Code
Within the PIN code menu, it is possible to set a four-digit number that can be used to
lock the menu feature of your OmniTest Twin Column. Please note once this has been
set you cannot access the menu without the PIN, so it is crucial that you keep a record
of this safe. If the PIN code has been set and then lost or is unknown, please contact your
local agent or Mecmesin Technical Support.
Settings: Information
This screen is used to display vital information relating to your MultiTest
-dV
and
connected devices. Here you can see software, hardware and firmware properties. If the
stand is operating in
dV
(u) mode you can also see the calibration date for ELS and the
number of overloads that have occurred for the current ELS.
Содержание MultiTest 0.5-dV
Страница 1: ...MultiTest dV Tension and Compression Test Stand Operating Manual 431 459 04 February 2019 ...
Страница 44: ...44 Mecmesin MultiTest dV Operating Manual MultiTest dV 1 0kN Dimensions Note The above drawing is not to scale ...
Страница 45: ...MultiTest dV Operating Manual Mecmesin 45 MultiTest dV 0 5kN Dimensions Note The above drawing is not to scale ...
Страница 46: ...46 Mecmesin MultiTest dV Operating Manual 11 Declaration of Conformity ...