Mecmesin
37
Limits
In a limit test, the crosshead cycles between the limit switches. To set your limit switches
loosen either screw, located on the rear of the test stand, and move the switch to the
required position.
A cycle starts when the crosshead is at the first limit switch and ends back at the same
limit switch. At the end of the test, the home button appears on the stand to enable the
crosshead to be moved back to the starting position.
Setting
Options
Cycle Count
0-9999
Up Speed
0-1200mm/min
Down Speed
0-1200mm/min
Start Direction
Choose whether the test direction is upward or downward.
Move to Start
Select if the test moves to the start position first.
Example
1.
Load your grips and sample,
2.
Set the upper limit switch at the desired upper displacement point,
3.
Set the lower limit switch at the desired lower displacement point,
4.
Set the desired start direction and whether to move to start,
5.
Configure the number of cycles,
6.
Execute the test. The stand moves between the two displacements for the configured
number of cycles and then stop.
Содержание MultiTest 0.5-dV
Страница 1: ...MultiTest dV Tension and Compression Test Stand Operating Manual 431 459 04 February 2019 ...
Страница 44: ...44 Mecmesin MultiTest dV Operating Manual MultiTest dV 1 0kN Dimensions Note The above drawing is not to scale ...
Страница 45: ...MultiTest dV Operating Manual Mecmesin 45 MultiTest dV 0 5kN Dimensions Note The above drawing is not to scale ...
Страница 46: ...46 Mecmesin MultiTest dV Operating Manual 11 Declaration of Conformity ...
Страница 51: ...MultiTest dV Operating Manual Mecmesin 51 14 Operator Notes ...