46 

 

MBW 473 Manual English V5.0 

At  equilibrium,  the  water  vapor  pressure  is  assumed  to  be  ho-

mogenous,  but  the  user  should  determine  this  by  experimenta-
tion according to the type of chamber being tested. Temperature 
probes should be positoned according to the need for validation. 
The  most  common  arrangement  is  four  probes  in  the  top 
corners,  four  in  the  bottom  corners  and  a  single  probe  in  the 
center. For larger chambers, more probes can be added so that 
the  entire  working  volume  is  measured.  This  allows  for  the  full 
spatial  volume  to  be  temperature  mapped.  By  combining  this 
temperature  data  with  the  measured  dew  point,  %RH  can  also 
be  mapped.  Software  such  as  that  supplied  with  the  MBW  T12 
temperature  system  can  be  used  to  display  and  record  the  dew  point,  temperature  and  calculated 
%RH values. Please contact MBW or your local supplier for more information. 

RH Generator Validation 

Transportable RH Generators provide a convenient  and fast 
way of calibrating RH instruments. Most use RH sensor con-
trol probes to measure and control the conditions in the gen-
erator

’s test chamber. The standard calibration uncertainty of 

the  RH  generator  is  limited  by  calibration  drift,  linearity  and 
temperature coefficients of the control probe. With the use of 
a  dew  point  mirror,  with  its  inherent  stability  and  precision, 
the uncertainty can be significantly improved. 

The 473-RP2 is ideal for this application. As shown, the RP2 
measuring  head  is  introduced  into  the  generator  chamber 
through a front access port. The measuring cell of the mirror 

is exposed to the moving air within the chamber   and measures the temperature and humidity of the 
air as it passes over the measuring head. As with climatic chambers, it is assumed that, at equilibrium, 
water vapour pressure is homogenous throughout the chamber volume. Provided that the temperature 
is stable and uniform, the %RH derived from the dew point and temperature measurement is suitable 
for use as a transfer standard for calibration of the generator

’s control probe or as a reference against 

which units under test can be compared. 

Temperature Effects 

 Most  RH  generators  are  also  able  to  control  temperature.  This 
means  that  RH  probes  can  be  calibrated  for  temperature  at  the 
same  time  as  RH,  and  RH  calibration  can  be  performed  at 
working  temperature.  As  RH  is  highly  temperature  dependent, 

temperature  effects  and  gradients  in  particular,  must  be 
considered. With the  RP2  probe  installed  into  the  RH  generator 
chamber,  the  Peltier  element  that  is  used  to  cool  the  mirror 
causes  heat  to  be  dissipated  from  the  body  of  the  measuring 
head. This can cause temperature gradients within the generator 
chamber,  so  these  effects  should  be  evaluated  and 
consequential uncertainty contributions applied to an uncertainty 
budget. 

 

Содержание 473-RP2

Страница 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...

Страница 2: ......

Страница 3: ...rized representatives may void all warranties Warranty may also be voided for misuse neglect accident corrosion and improper installation This Warranty is exclusive and in lieu of any and all other wa...

Страница 4: ...ii MBW 473 Manual English V5 0...

Страница 5: ...el 9 4 2 Touch Screen 12 4 3 Back Panel 13 4 4 Measuring Heads 18 4 5 Carrying Handle 23 5 SYSTEM CONFIGURATION 25 5 1 The Menu 25 5 2 Selection of indicated Parameters 26 5 3 Selection of Numeric or...

Страница 6: ...52 8 MAINTENANCE 55 8 1 Calibrate the Touch Screen 55 8 2 Mirror Cleaning 56 8 3 Exterior Cleaning 57 8 4 Peltier Cooling Test 58 8 5 Periodic Maintenance Checks 58 9 ERROR MESSAGES 59 9 1 Reversible...

Страница 7: ...housing or connecting disconnecting sensors Measuring head may get hot during operation Instruments built for the SH3 measuring head supply more power through the measuring head cable Therefore you m...

Страница 8: ...2 MBW 473 Manual English V5 0...

Страница 9: ...the calculation of relative humidity Optional Pressure Measurement Optionally the 473 can be ordered with internal pressure measurement capability This enables the user to measure barometric pressure...

Страница 10: ...4 MBW 473 Manual English V5 0...

Страница 11: ...easuring head and a 3 m cable to connect the probe direct ly to the back panel of the instrument o The SH2 is supplied with a 2 x 100 mm PRT and a 0 5 m cable to connect the probe to the measuring hea...

Страница 12: ...coefficients must be changed accordingly 3 2 Installing the Measuring Heads The heart of the 473 Dew Point Mirror Instrument is a highly sensitive and accurate measuring head Depending on your order t...

Страница 13: ...ternal Temperature Probe on page 16 Alternatively a fixed external temperature value may be entered via the touch screen Please follow the instructions in section 6 2 External Temperature You can test...

Страница 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...

Страница 15: ...ion on or off simply touch the desired key or object directly on the screen When the 473 is turned on the display will activate within a few seconds A sample display configura tion is shown below The...

Страница 16: ...below center The faster it disappears the lower the indication When the indicator is in the center the thickness of the dew or frost layer is neither growing nor evaporating and the layer on the mirr...

Страница 17: ...menu selection is circular Once you go past the last menu the first one will appears again and the process starts over You can use the key on the keypad to move backward through the menus Use the Ent...

Страница 18: ...h functions on or off touch the screen where appropriate with your finger or a stylus Never use sharp objects as damage may occur Before using the 473 for the first time or when several users operate...

Страница 19: ...MBW 473 Manual English V5 0 13 4 3 Back Panel Without Options With Barometric Pressure and Analog Output Options...

Страница 20: ...nnection The 473 can be ordered with different types of measuring heads the RP2 SH2 SH3 and SHX These measuring heads are connected to the back panel of the corresponding 473 display unit The 473 RP2...

Страница 21: ...he output signal is connected to a D A converter and then split into a voltage and a current signal There fore you may use either a volt or current meter to receive the analog signal The maximum volta...

Страница 22: ...ead 2 Use the supplied 4 pin 0 5 m cable to connect the probe to the measuring head 3 Use the supplied 5 pin 3 m cable to connect the probe directly to the back panel of the instrument The SHX is supp...

Страница 23: ...housing is located directly above pin 1 Pin Signal Position Description 1 Shield When viewing the solder tabs of a disassembled 5 pin LEMO connector pin 1 is usually identified with a full or partial...

Страница 24: ...5 m Cable 4 pin 6 19 pin Measuring Head Cable 2 m 7 Screw Cover 8 Optical Module 9 Guide Pin 10 Polished mirror SH2 Fan The fan of the SH2 measuring head is used to generate the necessary gas flow to...

Страница 25: ...4 pin 6 30 pin Measuring Head Cable 2 m 7 Screw Cover 8 Optical Module 9 Guide Pin 10 Polished Mirror SHX Fan The fan of the SHX measuring head is used to generate the necessary gas flow to the measu...

Страница 26: ...ad 1 Gas Inlet 6 mm Swagelok 2 Gas Outlet 6 mm Swagelok 3 15 pin Measuring Head Cable 2 m 4 2 pin Fan Cable 5 Measuring Head Fan 1 2 3 4 5 6 15 pin Measuring Head Cable Connector 7 Screw Cover 8 Optic...

Страница 27: ...One of the differences of the SH3 measuring head to the SH2 measuring head is that it does not have a sample gas flow fan The purpose of the fan installed on the SH3 is to cool the measuring head bas...

Страница 28: ...473 SH3 473 SHX Use only the corresponding measuring head for each display unit otherwise it will cause damage to the instrument The SHX version is equipped with a 30 pin measuring head connection ins...

Страница 29: ...Carrying Handle To adjust the position of the carrying han dle press the buttons on both sides Re lease the buttons when the handle is in the desired position Ensure that the handle has locked into pl...

Страница 30: ...24 MBW 473 Manual English V5 0...

Страница 31: ...red Staying in a specific menu will not affect the measurement List of available menus Parameter This menu is used to select the parameters displayed on the data lines Numeric Graphic This menu is use...

Страница 32: ...RH The ratio between the amount of water vapor in a sample and the maximum amount possi ble at that same temperature and pressure RH WMO The ratio between the amount of water vapor in a sample and th...

Страница 33: ...quire the measuring head to be connected such as Frost Point Dew Point Head Temp RH Humidity Volume Ratio Weight Ratio Vapor Pressure or Head Pressure please make sure all the relevant equipment is co...

Страница 34: ...key on the numeric key pad to confirm Press the Cancel button if you wish to abort all changes made in the dialog box Any values you enter will only be accepted by the system if they correspond with...

Страница 35: ...ge the time span the data that was sampled and stored at the old interval will be in crementally replaced by new data sampled at the new interval The time span indicated on the graph will always refle...

Страница 36: ...all other humidity measurements are then mathematically calculated from the frost point To change the Force Frost settings 1 Select the Control Setup menu by pressing the dark gray menu selection key...

Страница 37: ...t it is also possible that dew and frost exist concurrently on the mirror which results in a non stable value reading somewhere between the dew and frost point Therefore the phase of the condensate mu...

Страница 38: ...that holds the number you wish to change Each time you touch the field the last digit entered will be deleted 6 Press the OK button in the dialog box or press Enter on the keypad to confirm the new va...

Страница 39: ...ld before allowing the next dew or frost point measurement A heating time of 0 means that the instrument will resume dew or frost point measurement immediately upon reaching the mirror check temperatu...

Страница 40: ...ish to change Each time you touch the field the last digit entered will be deleted Press the Ok button in the dialog box or press Enter on the keypad to confirm the new value Press Cancel to leave it...

Страница 41: ...ata retrieved via RS 232 will always be in SI units regardless of the units chosen for display Also note that settings within dialog boxes used for changing system parameters are entered and displayed...

Страница 42: ...pointing arrows Note that each of the upper keys correspond to the data lines they point to 2 Press the arrow key of the data line you wish to change Note that the foreground color of the data line w...

Страница 43: ...read the three following subsections 4 Follow the same procedure for the sec ond or any subsequent analog outputs as needed Selection of Parameter to Track In the analog configuration window enter the...

Страница 44: ...1 VDC on the analog out put to represent the 0 100 RH Enter 0 into the field next to the Min Voltage and 1 into the field next to Max Voltage Example 2 To keep things simple we will take the same Para...

Страница 45: ...suspect that there may be an error with the instrument The mirror must be visible to perform the Ice Test If you have an SH2 SH3 or SHX measuring head remove the measuring head cover prior to the Ice...

Страница 46: ...ase transition When you observe the phase transition on the mirror press the Ok button The mirror temperature is measured at that moment and a dialog box appears with the test results If the measured...

Страница 47: ...wn in seconds until the test ends The duration of the test is two minutes TH Temperature of the measuring head TM Actual temperature of the mirror TMdelta Delta between head temperature and mirror tem...

Страница 48: ...ay menu selection key until Diagnostics Functions ap pears 2 Touch the RS 232 button The RS 232 Configuration window will open 3 To change the baud rate value touch the gray Change To field The field...

Страница 49: ...Set Up and Operation 6 1 Measurement Set Up Back Panel Connections The following picture shows the back panel of a fully connected 473 with RS 232 cable 19 pin meas uring head cable power cable and th...

Страница 50: ...rature Enter the desired temperature in degrees Celsi us 21 C in this example and activate the Use Fixed Ext Temp option by touching the button When the indicator square turns green the fixed external...

Страница 51: ...properties of the measuring head and any items in its environment and their possible effects The following application examples are included to show some of the most common applications of the 473 In...

Страница 52: ...inherent stability and precision the uncertainty can be significantly improved The 473 RP2 is ideal for this application As shown the RP2 measuring head is introduced into the generator chamber throu...

Страница 53: ...sufficient according to IEC 60068 When comparing for example RH probes to the RP2 probe mounted in an RH generator chamber it is reccomended that there is at least 10mm of clearance between the probes...

Страница 54: ...rature measurement system may not be practical for site engineers as they need to run calibrations quickly to minimize down time Examples of this situation include calibration of instru ments used in...

Страница 55: ...t connector rather than a 9 pin connector you will also need a 25 pin to 9 pin port adapter Both the 9 pin RS 232 extender cable and the 25 pin to 9 pin port adapter are commonly available at most com...

Страница 56: ...3 will be shown in this font General Use All commands require a question mark to indicate you are requesting data When requesting data from the 473 follow the command with the question mark character...

Страница 57: ...tation Receiving a number as 12 34 is the same as receiving it as 1234e 2 or as 1 234e1 Depending on the value of numerical responses the 473 sends out it may send the numbers in either standard or sc...

Страница 58: ...fic Humidity g kg VP Vapor Pressure Pa P Head Pressure Pa Tx External Temperature C Tm Mirror Temperature C Th Head Temperature C Om Mirror PRT Resistance Ohms Ox External PRT Resistance Ohms System I...

Страница 59: ...tivates AMC 0 disables AMC AMC cycleTime n AMC cycle time in minutes AMC heatTime n AMC heating hold time in minutes AMC temp n Target temperature during AMC C AMC dispHold i 1 enables frost dew point...

Страница 60: ...converter is calibrated Automatically sets to Ana1 hold 1 If calibration done Ana1 hold must be sent 0 Ana1 hold i 1 calibration mode Analog output is set with the command Ana1 output n 0 normal mode...

Страница 61: ...key on the numerical keypad for 3 to 4 seconds You will hear two short beeps and the key in the upper right corner will turn yellow With the tip of your finger press the center of the yellow key in t...

Страница 62: ...red knurled knob To remove it simply turn it counterclockwise It re quires approximately three full turns to completely unscrew it Once the screw cover has been removed remove the black optical module...

Страница 63: ...opto electronic ele ments and gold contacts The other half con tains the mirror temperature sensor and some mating gold contacts Avoid touching the mirror and gold contacts with your fingers to preven...

Страница 64: ...ou can check this with the Peltier Cooling Test PCT function Please follow the instructions in section Peltier Cooling Test on page 41 8 5 Periodic Maintenance Checks Apart from periodic mirror cleani...

Страница 65: ...Chamber temperature may be too high Try again after several minutes if it fails again contact MBW service 9 2 Irreversible Error Messages Error Message Explanation Possible Reasons and Solutions ERROR...

Страница 66: ...lay unit does not recog nize a fan No fan is mounted to the measur ing head There is a cable break between the display unit and the fan Con tact MBW service Fan supply failure The fan driver in the di...

Страница 67: ...ation certificate RS 232 5 7 LCD with color touch screen Platinum Resistance Thermometer IEC 60751 2008 2 5 m English Factory calibration 5 points FP DP 3 points temperature Optional Internal barometr...

Страница 68: ...62 MBW 473 Manual English V5 0...

Страница 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...

Страница 70: ...64 MBW 473 Manual English V5 0 Front View Back View...

Страница 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...

Страница 72: ...66 MBW 473 Manual English V5 0 SHX SH3...

Страница 73: ...MBW 473 Manual English V5 0 67...

Страница 74: ...68 MBW 473 Manual English V5 0...

Страница 75: ...y was configured incorrectly and I would like to go back to factory default Solution Press and hold key number 7 or 8 on the keyboard for 3 4 seconds This enables you to restore the factory default se...

Отзывы: