44
MBW 473 Manual English V5.0
6.2 External Temperature
Set Fixed External Temperature
If you do not use an external temperature sen-
sor, you have the possibility to enter a fixed
external temperature. Set the parameter so that
External Temp
shows on one data line (see
section 0). If no temperature sensor is installed,
the data line will not show any reading. Touch
the
External Temp
data line on the screen. A
window (see picture below) will appear where
you can enter the desired temperature.
Enter the desired temperature in degrees Celsi-
us (21 °C in this example) and activate the
Use
Fixed Ext Temp
option by touching the button.
When the indicator square turns green, the
fixed external temperature is active.
The External Temp data line (and the graph, if
shown) will now say
“
(fixed)
”
. As shown in the
image on the right, the external temperature will
immediately change to the temperature that was
entered, and will remain there until it is changed
again.
If you want to start using an external tempera-
ture probe, go to
the Fixed External Tempera-
ture
window and press the
Use Fixed Ext
Temp
to disable the fixed external temperature
function. The green square will turn grey.
Содержание 473-RP2
Страница 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...
Страница 2: ......
Страница 4: ...ii MBW 473 Manual English V5 0...
Страница 8: ...2 MBW 473 Manual English V5 0...
Страница 10: ...4 MBW 473 Manual English V5 0...
Страница 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...
Страница 30: ...24 MBW 473 Manual English V5 0...
Страница 68: ...62 MBW 473 Manual English V5 0...
Страница 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...
Страница 70: ...64 MBW 473 Manual English V5 0 Front View Back View...
Страница 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...
Страница 72: ...66 MBW 473 Manual English V5 0 SHX SH3...
Страница 73: ...MBW 473 Manual English V5 0 67...
Страница 74: ...68 MBW 473 Manual English V5 0...