54
MBW 473 Manual English V5.0
Mirror Temperature Calibration Coefficients
Syntax
Function
MirrorTempCal.r0[=n][?]
Callendar VanDusen R0 coefficient, ~ 100.00
MirrorTempCal.a[=n][?]
Callendar VanDusen A coefficient, ~ 3.9083e-3
MirrorTempCal.b[=n][?]
Callendar VanDusen B coefficient, ~
–5.775e-7
MirrorTempCal.c[=n][?]
Callendar VanDusen C coefficient, ~ -4.183e-12
MirrorTempCal.avg[=n][?]
Amount of averaging applied to measurement, ~40
MirrorTempCal.avgBand[=i][?]
Only average when measurements remain within
this band, ~ 0.2 °C
SaveCal=473
Save these and other calibration parameters
Analog Output (optional)
Syntax
Function
Ana1.param[=i][?]
Parameter configuration for analogue output 1
Ana1.paramMin[=n][?]
Min. Value
Ana1.paramMax[=n][?]
Max. Value
Ana1.OutMin[=n][?]
Min. Voltage (-10V)
Ana1.OutMax[=n][?]
Max. Voltage (+10V)
Ana1.calGain[=n][?]
Gain scaling factor for DAC
Ana1.calZero[=n][?]
Zero scaling factor for DAC
Ana1.output=n
Requires to set the analog output when the DAC
converter is calibrated. Automatically sets to
Ana1.hold = 1 If calibration done, Ana1.hold must
be sent = 0.
Ana1.hold[=i][?]
1 calibration mode. Analog output is set with the
command Ana1.output = n
0 normal mode. Analog output the parameters
monitored set with Ana1.param = i
SaveCfg=473
Saves all configurations to 473
Ana2.param[=i][?]
Parameter configuration for analogue output 2
Ana2.paramMin[=n][?]
Min. Value
Ana2.paramMax[=n][?]
Max. Value
Ana2.OutMin[=n][?]
Min. Voltage (-10V)
Ana2.OutMax[=n][?]
Max. Voltage (+10V)
Ana2.calGain[=n][?]
Gain scaling factor for DAC
Ana2.calZero[=n][?]
Zero scaling factor for DAC
Ana2.output=n
Requires to set the analog output when the DAC
converter is calibrated. Automatically sets to
Ana1.hold = 1 If calibration done, Ana1.hold must
be sent = 0.
Ana2.hold[=i][?]
1 calibration mode. Analog output is set with the
command Ana1.output = n
0 normal mode. Analog output the parameters
monitored set with Ana1.param = i
SaveCfg=473
Saves all configurations to 473
Communication Settings
Syntax
Function
Baud[=n][?]
Set Baud Rate: 1200, 2400, 4800, 9600, 19200,
38400, 57600, 115200. (Default rate is 9600)
SaveCfg=473
Saves all configurations to 473
Содержание 473-RP2
Страница 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...
Страница 2: ......
Страница 4: ...ii MBW 473 Manual English V5 0...
Страница 8: ...2 MBW 473 Manual English V5 0...
Страница 10: ...4 MBW 473 Manual English V5 0...
Страница 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...
Страница 30: ...24 MBW 473 Manual English V5 0...
Страница 68: ...62 MBW 473 Manual English V5 0...
Страница 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...
Страница 70: ...64 MBW 473 Manual English V5 0 Front View Back View...
Страница 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...
Страница 72: ...66 MBW 473 Manual English V5 0 SHX SH3...
Страница 73: ...MBW 473 Manual English V5 0 67...
Страница 74: ...68 MBW 473 Manual English V5 0...