![MBW 473-RP2 Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/mbw/473-rp2/473-rp2_operation-and-maintenance-manual_1752851066.webp)
60
MBW 473 Manual English V5.0
9.3 Fan failure SHX
Error Message
Explanation
Possible Reasons and Solutions
Fan not found
The display unit does not recog-
nize a fan
-
No fan is mounted to the measur-
ing head
-
There is a cable break between
the display unit and the fan. Con-
tact MBW service
Fan supply failure
The fan driver in the display unit
detects that the supply voltage
of the fan is missing/is too low
-
A blown fuse on the fan driver
PCB. Contact MBW service
Содержание 473-RP2